Long-term outcome of perimembranous VSD closure using the Nit-Occlud® Le VSD coil system.


ABSTRACT:

Objective

This study presents data from the admission trial to show the feasibility, safety and effectiveness of the Nit-Occlud® Lê VSD in the treatment of perimembranous ventricular septal defects with an aneurysmal configuration and a diameter up to 8 mm.

Background

The majority of ventricular septal defects (VSD) are still closed surgically, while a less invasive transcatheter treatment by closure devices is available. Device-based closure is reported to be associated with the risk of complete atrio-ventricular block, especially with double-disc devices in perimembranous defects.

Methods

In six tertiary centers in Germany and Israel, an interventional closure of a periembranous VSD was attempted in 88 patients using the Nit-Occlud® Lê VSD.

Results

The interventional VSD closure was performed in 85 patients. Patients had a median age of 8.0 (2-65) years and a median body weight of 26.7 (10-109) kg. A complete closure of the defects was achieved in 85.4% 2 weeks after device implantation, in 88.9% after three months and in 98.6% at the 5-year follow-up. There was no incidence of death during the study nor did any patient suffer of permanent atrio-ventricular block of higher degree. Serious adverse events, by definition, are potentially life-threatening or require surgery to correct, while major serious events require medical or transcatheter intervention to correct. The study results exhibit a serious adverse event rate of 3.5% (3/85 patients) and a major adverse event rate of 5.9% (5/85 patients).

Conclusion

The Nit-Occlud® Lê VSD coil offers the possibility of an effective and safe approach in patients with aneurysmal perimembranous ventricular septal defects.

SUBMITTER: Kozlik-Feldmann R 

PROVIDER: S-EPMC7906931 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Long-term outcome of perimembranous VSD closure using the Nit-Occlud® Lê VSD coil system.

Kozlik-Feldmann Rainer R   Lorber Avraham A   Sievert Horst H   Ewert Peter P   Jux Christian C   Müller Götz C GC   Dalla Pozza Robert R   Yigitbasi Mustafa M   Schranz Dietmar D   Lindinger Angelika A   Galal Omar O   Meinertz Thomas T  

Clinical research in cardiology : official journal of the German Cardiac Society 20201031 3


<h4>Objective</h4>This study presents data from the admission trial to show the feasibility, safety and effectiveness of the Nit-Occlud® Lê VSD in the treatment of perimembranous ventricular septal defects with an aneurysmal configuration and a diameter up to 8 mm.<h4>Background</h4>The majority of ventricular septal defects (VSD) are still closed surgically, while a less invasive transcatheter treatment by closure devices is available. Device-based closure is reported to be associated with the  ...[more]

Similar Datasets

| S-EPMC5545148 | biostudies-other
| S-EPMC6827482 | biostudies-literature
| S-EPMC9640380 | biostudies-literature
| S-EPMC2938468 | biostudies-literature
| S-EPMC7965457 | biostudies-literature
| S-EPMC6794751 | biostudies-literature
| S-EPMC10292574 | biostudies-literature
| S-EPMC7293941 | biostudies-literature
| S-EPMC5040263 | biostudies-literature
| S-EPMC4699954 | biostudies-literature