Au-Ag assembled on silica nanoprobes for visual semiquantitative detection of prostate-specific antigen.


ABSTRACT:

Background

Blood prostate-specific antigen (PSA) levels are widely used as diagnostic biomarkers for prostate cancer. Lateral-flow immunoassay (LFIA)-based PSA detection can overcome the limitations associated with other methods. LFIAbased PSA detection in clinical samples enables prognosis and early diagnosis owing to the use of high-performance signal reporters.

Results

Here, a semiquantitative LFIA platform for PSA detection in blood was developed using Au-Ag nanoparticles (NPs) assembled on silica NPs (SiO2@Au-Ag NPs) that served as signal reporters. Synthesized SiO2@Au-Ag NPs exhibited a high absorbance at a wide wavelength range (400-800 nm), with a high scattering on nitrocellulose membrane test strips. In LFIA, the color intensity of the test line on the test strip differed depending on the PSA concentration (0.30-10.00 ng/mL), and bands for the test line on the test strip could be used as a standard. When clinical samples were assessed using this LFIA, a visual test line with particular color intensity observed on the test strip enabled the early diagnosis and prognosis of patients with prostate cancer based on PSA detection. In addition, the relative standard deviation of reproducibility was 1.41%, indicating high reproducibility, and the signal reporter showed good stability for 10 days.

Conclusion

These characteristics of the signal reporter demonstrated the reliability of the LFIA platform for PSA detection, suggesting potential applications in clinical sample analysis.

SUBMITTER: Kim HM 

PROVIDER: S-EPMC7953718 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Au-Ag assembled on silica nanoprobes for visual semiquantitative detection of prostate-specific antigen.

Kim Hyung-Mo HM   Kim Jaehi J   An Jaehyun J   Bock Sungje S   Pham Xuan-Hung XH   Huynh Kim-Hung KH   Choi Yoonsik Y   Hahm Eunil E   Song Hobeom H   Kim Jung-Won JW   Rho Won-Yeop WY   Jeong Dae Hong DH   Lee Ho-Young HY   Lee Sangchul S   Jun Bong-Hyun BH  

Journal of nanobiotechnology 20210312 1


<h4>Background</h4>Blood prostate-specific antigen (PSA) levels are widely used as diagnostic biomarkers for prostate cancer. Lateral-flow immunoassay (LFIA)-based PSA detection can overcome the limitations associated with other methods. LFIAbased PSA detection in clinical samples enables prognosis and early diagnosis owing to the use of high-performance signal reporters.<h4>Results</h4>Here, a semiquantitative LFIA platform for PSA detection in blood was developed using Au-Ag nanoparticles (NPs  ...[more]

Similar Datasets

| S-EPMC8232291 | biostudies-literature
| S-EPMC8145968 | biostudies-literature
| S-EPMC5477113 | biostudies-literature
| S-EPMC6099385 | biostudies-literature
| S-EPMC6445958 | biostudies-literature
| S-EPMC5453564 | biostudies-literature
| S-EPMC6751705 | biostudies-literature
| S-EPMC5908853 | biostudies-literature
| S-EPMC8746978 | biostudies-literature
| S-EPMC4210868 | biostudies-literature