Cerebral phosphoester signals measured by 31P magnetic resonance spectroscopy at 3 and 7 Tesla.
Ontology highlight
ABSTRACT: Abnormal cell membrane metabolism is associated with many neuropsychiatric disorders. Free phosphomonoesters and phosphodiesters, which can be detected by in vivo 31P magnetic resonance spectroscopy (MRS), are important cell membrane building blocks. However, the quantification of phosphoesters has been highly controversial even in healthy individuals due to overlapping signals from macromolecule membrane phospholipids (MP). In this study, high signal-to-noise ratio (SNR) cerebral 31P MRS spectra were acquired from healthy volunteers at both 3 and 7 Tesla. Our results indicated that, with minimal spectral interference from MP, the [phosphocreatine (PCr)]/[phosphocholine (PC) + glycerophosphocholine (GPC)] ratio measured at 7 Tesla agreed with its value expected from biochemical constraints. In contrast, the 3 Tesla [PCr]/[PC+GPC] ratio obtained using standard spectral fitting procedures was markedly smaller than the 7 Tesla ratio and than the expected value. The analysis suggests that the commonly used spectral model for MP may fail to capture its complex spectral features at 3 Tesla, and that additional prior knowledge is necessary to reliably quantify the phosphoester signals at low magnetic field strengths when spectral overlapping is significant.