Unknown

Dataset Information

0

Optimization of adiabatic pulses for pulsed arterial spin labeling at 7 tesla: Comparison with pseudo-continuous arterial spin labeling.


ABSTRACT:

Purpose

To optimize and evaluate adiabatic pulses for pulsed arterial spin labeling at ultrahigh field 7 tesla.

Methods

Four common adiabatic inversion pulses, including hyperbolic secant, wideband uniform rate smooth truncation, frequency offset corrected inversion, and time-resampled frequency offset corrected inversion pulses, were optimized based on a custom-defined loss function that included labeling efficiency and inversion band uniformity. The optimized pulses were implemented in flow-sensitive alternating inversion recovery sequences and tested on phantom and 11 healthy volunteers with 2 constraints: 1) specific absorption rate normalized; and 2) equal peak RF amplitude, respectively. A pseudo-continuous arterial spin labeling sequence was implemented for comparison. Quantitative metrics such as perfusion and relative labeling efficiency versus residual tissue signal were calculated.

Results

Among the 4 pulses, the wideband uniform rate smooth truncation pulse yielded the lowest loss in simulation and achieved a good balance between labeling efficiency and residual tissue signal from both phantom and in vivo experiments. Wideband uniform rate smooth truncation-pulsed arterial spin labeling showed significantly higher relative labeling efficiency compared to the other sequences (P < .01), whereas the perfusion signal was increased by 40% when the highest B1+ amplitude was used. The 4 pulsed arterial spin labeling sequences yielded comparable perfusion signals compared to pseudo-continuous arterial spin labeling but with less than half the specific absorption rate.

Conclusion

Optimized wideband uniform rate smooth truncation pulse with the highest B1+ amplitude allowed was recommended for 7 tesla pulsed arterial spin labeling.

SUBMITTER: Wang K 

PROVIDER: S-EPMC8351166 | biostudies-literature | 2021 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Optimization of adiabatic pulses for pulsed arterial spin labeling at 7 tesla: Comparison with pseudo-continuous arterial spin labeling.

Wang Kai K   Shao Xingfeng X   Yan Lirong L   Ma Samantha J SJ   Jin Jin J   Wang Danny J J DJJ  

Magnetic resonance in medicine 20210111 6


<h4>Purpose</h4>To optimize and evaluate adiabatic pulses for pulsed arterial spin labeling at ultrahigh field 7 tesla.<h4>Methods</h4>Four common adiabatic inversion pulses, including hyperbolic secant, wideband uniform rate smooth truncation, frequency offset corrected inversion, and time-resampled frequency offset corrected inversion pulses, were optimized based on a custom-defined loss function that included labeling efficiency and inversion band uniformity. The optimized pulses were impleme  ...[more]

Similar Datasets

| S-EPMC4215516 | biostudies-literature
| S-EPMC8616784 | biostudies-literature
| S-EPMC8810716 | biostudies-literature
| S-EPMC3170937 | biostudies-other
| S-EPMC3121564 | biostudies-literature
| S-EPMC10528743 | biostudies-literature
| S-EPMC10168494 | biostudies-literature
| S-EPMC10091734 | biostudies-literature
| S-EPMC6899553 | biostudies-literature