Optimization of pseudo-continuous arterial spin labeling at 7T with parallel transmission B1 shimming.


ABSTRACT:

Purpose

To optimize pseudo-continuous arterial spin labeling (pCASL) for 7 T, and to further improve the labeling efficiency with parallel RF transmission transmit B1 ( B1+ ) shimming.

Methods

pCASL parameters were optimized based on B1+/B0 field distributions at 7 T with simulation. To increase labeling efficiency, the B1+ amplitude at inflowing arteries was increased with parallel RF transmission B1+ shimming. The "indv-shim" with shimming weights calculated for each individual subject, and the "univ-shim" with universal weights calculated on a group of 12 subjects, were compared with circular polarized (CP) shim. The optimized pCASL sequences with three B1+ shimming modes (indv-shim, univ-shim, and CP-shim) were evaluated in 6 subjects who underwent two repeated scans 24 hours apart, along with a pulsed ASL sequence. Quantitative metrics including mean B1+ amplitude, perfusion, and intraclass correlation coefficient were calculated. The optimized 7T pCASL was compared with standard 3T pCASL on 5 subjects, using spatial SNR and temporal SNR.

Results

The optimal pCASL parameter set (RF duration/gap = 300/250 us, Gave=0.6mT/m,gRatio=10 ) achieved robust perfusion measurement in the presence of B1+/B0 inhomogeneities. Both indv-shim and univ-shim significantly increased B1+ amplitude compared with CP-shim in simulation and in vivo experiment (P < .01). Compared with CP-shim, perfusion signal was increased by 9.5% with indv-shim (P < .05) and by 5.3% with univ-shim (P = .35). All three pCASL sequences achieved fair to good repeatability (intraclass correlation coefficient ≥ 0.5). Compared with 3T pCASL, the optimized 7T pCASL achieved 78.3% higher spatial SNR and 200% higher temporal SNR.

Conclusion

The optimized pCASL achieved robust perfusion imaging at 7 T, while both indv-shim and univ-shim further increased labeling efficiency.

SUBMITTER: Wang K 

PROVIDER: S-EPMC8616784 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Optimization of pseudo-continuous arterial spin labeling at 7T with parallel transmission B1 shimming.

Wang Kai K   Ma Samantha J SJ   Shao Xingfeng X   Zhao Chenyang C   Shou Qinyang Q   Yan Lirong L   Wang Danny J J DJJ  

Magnetic resonance in medicine 20210824 1


<h4>Purpose</h4>To optimize pseudo-continuous arterial spin labeling (pCASL) for 7 T, and to further improve the labeling efficiency with parallel RF transmission transmit B<sub>1</sub> ( B1+ ) shimming.<h4>Methods</h4>pCASL parameters were optimized based on B1+/B0 field distributions at 7 T with simulation. To increase labeling efficiency, the B1+ amplitude at inflowing arteries was increased with parallel RF transmission B1+ shimming. The "indv-shim" with shimming weights calculated for each  ...[more]

Similar Datasets

| S-EPMC8810716 | biostudies-literature
| S-EPMC10528743 | biostudies-literature
| S-EPMC10168494 | biostudies-literature
| S-EPMC8351166 | biostudies-literature
| S-EPMC3121564 | biostudies-literature
| S-EPMC4215516 | biostudies-literature
| S-EPMC10091734 | biostudies-literature
| S-EPMC6899553 | biostudies-literature
| S-EPMC7100033 | biostudies-literature