Pelvic bone CT: can tin-filtered ultra-low-dose CT and virtual radiographs be used as alternative for standard CT and digital radiographs?


ABSTRACT:

Objectives

To compare ultra-low-dose CT (ULD-CT) of the osseous pelvis with tin filtration to standard clinical CT (CT), and to assess the quality of computed virtual pelvic radiographs (VRs).

Methods

CT protocols were optimized in a phantom and three pelvic cadavers. Thirty prospectively included patients received both standard CT (automated tube voltage selection and current modulation) and tin-filtered ULD-CT of the pelvis (Sn140kV/50mAs). VRs of ULD-CT data were computed using an adapted cone beam-based projection algorithm and were compared to digital radiographs (DRs) of the pelvis. CT and DR dose parameters and quantitative and qualitative measures (1 = worst, 4 = best) were compared. CT and ULD-CT were assessed for osseous pathologies.

Results

Dose reduction of ULD-CT was 84% compared to CT, with a median effective dose of 0.38 mSv (quartile 1-3: 0.37-0.4 mSv) versus 2.31 mSv (1.82-3.58 mSv; p < .001), respectively. Mean dose of DR was 0.37 mSv (± 0.14 mSv). The median signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of bone were significantly higher for CT (64.3 and 21.5, respectively) compared to ULD-CT (50.4 and 18.8; p ≤ .01), while ULD-CT was significantly more dose efficient (figure of merit (FOM) 927.6) than CT (FOM 167.6; p < .001). Both CT and ULD-CT were of good image quality with excellent depiction of anatomy, with a median score of 4 (4-4) for both methods (p = .1). Agreement was perfect between both methods regarding the prevalence of assessed osseous pathologies (p > .99). VRs were successfully calculated and were equivalent to DRs.

Conclusion

Tin-filtered ULD-CT of the pelvis at a dose equivalent to standard radiographs is adequate for assessing bone anatomy and osseous pathologies and had a markedly superior dose efficiency than standard CT.

Key points

• Ultra-low-dose pelvic CT with tin filtration (0.38 mSv) can be performed at a dose of digital radiographs (0.37 mSv), with a dose reduction of 84% compared to standard CT (2.31 mSv). • Tin-filtered ultra-low-dose CT had lower SNR and CNR and higher image noise than standard CT, but showed clear depiction of anatomy and accurate detection of osseous pathologies. • Virtual pelvic radiographs were successfully calculated from ultra-low-dose CT data and were equivalent to digital radiographs.

SUBMITTER: Stern C 

PROVIDER: S-EPMC8379132 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC6439109 | biostudies-literature
| S-EPMC9701766 | biostudies-literature
| S-EPMC7001933 | biostudies-literature
| S-EPMC2504670 | biostudies-other
| S-EPMC2504653 | biostudies-other
| S-EPMC9883314 | biostudies-literature
| S-EPMC4039784 | biostudies-literature
| S-EPMC5708766 | biostudies-literature
| S-EPMC6701577 | biostudies-literature
| S-EPMC6443697 | biostudies-literature