Determining Particle Size and Position in a Coplanar Electrode Setup Using Measured Opacity for Microfluidic Cytometry.


ABSTRACT: Microfluidic impedance flow cytometers enable high-throughput, non-invasive, and label-free detection of single-cells. Cytometers with coplanar electrodes are easy and cheap to fabricate, but are sensitive to positional differences of passing particles, owing to the inhomogeneous electric field. We present a novel particle height compensation method, which employs the dependence of measured electrical opacity on particle height. The measured electrical opacity correlates with the particle height as a result of the constant electrical double layer series capacitance of the electrodes. As an alternative to existing compensation methods, we use only two coplanar electrodes and multi-frequency analysis to determine the particle size of a mixture of 5, 6, and 7 µm polystyrene beads with an accuracy (CV) of 5.8%, 4.0%, and 2.9%, respectively. Additionally, we can predict the bead height with an accuracy of 1.5 µm (8% of channel height) using the measured opacity and we demonstrate its application in flow cytometry with yeast. The use of only two electrodes is of special interest for simplified, easy-to-use chips with a minimum amount of instrumentation and of limited size.

SUBMITTER: de Bruijn DS 

PROVIDER: S-EPMC8533872 | biostudies-literature | 2021 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Determining Particle Size and Position in a Coplanar Electrode Setup Using Measured Opacity for Microfluidic Cytometry.

de Bruijn Douwe S DS   Jorissen Koen F A KFA   Olthuis Wouter W   van den Berg Albert A  

Biosensors 20210923 10


Microfluidic impedance flow cytometers enable high-throughput, non-invasive, and label-free detection of single-cells. Cytometers with coplanar electrodes are easy and cheap to fabricate, but are sensitive to positional differences of passing particles, owing to the inhomogeneous electric field. We present a novel particle height compensation method, which employs the dependence of measured electrical opacity on particle height. The measured electrical opacity correlates with the particle height  ...[more]

Similar Datasets

| S-EPMC8590873 | biostudies-literature
| S-EPMC11244668 | biostudies-literature
| S-EPMC7007699 | biostudies-literature
| S-EPMC9352041 | biostudies-literature
| S-EPMC2707935 | biostudies-literature
| S-EPMC11487660 | biostudies-literature
| S-EPMC5266633 | biostudies-literature
| S-EPMC3559333 | biostudies-literature
| S-EPMC11509203 | biostudies-literature
| S-EPMC3145043 | biostudies-literature