Unknown

Dataset Information

0

Optical property recovery with spatially-resolved diffuse reflectance at short source-detector separations using a compact fiber-optic probe.


ABSTRACT: We describe a compact fiber-optic probe (2 mm outside diameter) that utilizes spatially-resolved diffuse reflectance for tissue optical property recovery. Validation was performed in phantoms containing Intralipid 20% as scatterer, and methylene blue (MB), MnTPPS, and/or India ink as absorbers. Over a range of conditions, the reduced scattering coefficient was recovered with a root mean square error (RMSE) of 0.86-2.7 cm-1 (average error = 3.8%). MB concentration was recovered with RMSE = 0.26-0.52 µM (average error = 15.0%), which did not vary with inclusion of MnTPPS (p=0.65). This system will be utilized to determine optical properties in human abscesses, in order to generate treatment plans for photodynamic therapy.

SUBMITTER: Bridger KG 

PROVIDER: S-EPMC8713658 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC8692235 | biostudies-literature
| S-EPMC5266677 | biostudies-literature
| S-EPMC8235001 | biostudies-literature
| S-EPMC2673342 | biostudies-literature
| S-EPMC5714651 | biostudies-literature
| S-EPMC3420182 | biostudies-literature
| S-EPMC3041242 | biostudies-other
| S-EPMC4815871 | biostudies-other
| S-EPMC5504667 | biostudies-literature
| S-EPMC7710487 | biostudies-literature