ABSTRACT: Background
Stereotactic body radiotherapy (SBRT) has been shown to be a promising therapy for unresectable pancreatic tumors. However, intrafraction motion, caused by respiratory motion and organ drift, is one of the main concerns for efficient dose delivery in ungated upper abdominal radiotherapy. The aim of this study was to analyze the intrafraction gross tumor volume (GTV) motion in a clinical cohort.Materials and methods
We included 13 patients that underwent online adaptive magnetic resonance (MR)-guided SBRT for malignancies in the pancreatic region (5 × 8 Gy). An abdominal corset was fitted in order to reduce the abdominal respiratory motion. Coronal and sagittal cine magnetic resonance images of the tumor region were made at 2 Hz during the entire beam-on time of each fraction. We used deformable image registration to obtain GTV motion profiles in all three directions, which were subsequently high-pass and low-pass filtered to isolate the motion caused by respiratory motion and baseline drift, respectively.Results
The mean (SD) respiratory amplitudes were 4.2 (1.9) mm cranio-caudal (CC), 2.3 (1.1) mm ventral-dorsal (AP) and 1.4 (0.6) mm left-right (LR), with low variability within patients. The mean (SD) maximum baseline drifts were 1.2 (1.1) mm CC, 0.5 (0.4) mm AP and 0.5 (0.3) mm LR. The mean (SD) minimum baseline drifts were -0.7 (0.5) mm CC, -0.6 (0.5) mm AP and -0.5 (0.4) mm LR.Conclusion
Overall tumor motion during treatment was small and interfractionally stable. These findings show that high-precision ungated MR-guided SBRT is feasible with an abdominal corset.