Intrafraction motion during radiotherapy of breast tumor, breast tumor bed, and individual axillary lymph nodes on cine magnetic resonance imaging


ABSTRACT: Graphical abstract Highlights • Intrafraction motion of the breast and individual axillary lymph nodes was studied.• Displacements were investigated using cine magnetic resonance imaging.• Motion was separated into breathing and drift components.• Medians of the maximum displacements were small, <3 mm for breast and lymph nodes.• Intrafraction motion of the tumor (bed) was less in prone than in supine position.

Background and purpose

In (ultra-)hypofractionation, the contribution of intrafraction motion to treatment accuracy becomes increasingly important. Our purpose was to evaluate intrafraction motion and resulting geometric uncertainties for breast tumor (bed) and individual axillary lymph nodes, and to compare prone and supine position for the breast tumor (bed).

Materials and methods

During 1–3 min of free breathing, we acquired transverse/sagittal interleaved 1.5 T cine magnetic resonance imaging (MRI) of the breast tumor (bed) in prone and supine position and coronal/sagittal cine MRI of individual axillary lymph nodes in supine position. A total of 31 prone and 23 supine breast cine MRI (in 23 women) and 52 lymph node cine MRI (in 24 women) were included. Maximum displacement, breathing amplitude, and drift were analyzed using deformable image registration. Geometric uncertainties were calculated for all displacements and for breathing motion only.

Results

Median maximum displacements (range over the three orthogonal orientations) were 1.1–1.5 mm for the breast tumor (bed) in prone and 1.8–3.0 mm in supine position, and 2.2–2.4 mm for lymph nodes. Maximum displacements were significantly smaller in prone than in supine position, mainly due to smaller breathing amplitude: 0.6–0.9 mm in prone vs. 0.9–1.4 mm in supine. Systematic and random uncertainties were 0.1–0.4 mm in prone position and 0.2–0.8 mm in supine position for the tumor (bed), and 0.4–0.6 mm for the lymph nodes.

Conclusion

Intrafraction motion of breast tumor (bed) and individual lymph nodes was small. Motion of the tumor (bed) was smaller in prone than in supine position.