Real-time CMR guidance for intracardiac and great vessel pressure mapping in patients with congenital heart disease using an MR conditional guidewire-results of 25 patients.


ABSTRACT:

Background

The aim of this study was to test a CE-certified MR-conditional guidewire to facilitate blood pressure measurement in cardiovascular magnetic resonance (CMR) using fluid-filled catheters in patients with congenital heart disease (CHD). The main purpose was to determine procedural success in a post market clinical follow-up (PMCF) for routine procedure in a diagnostic and interventional workflow. Real-time CMR provides high quality imaging without the risk of exposing the patient to X-rays, especially for patients with irregular heart anatomy and patients who are susceptible to radiation and iodinated contrast media. To date, the assessment of blood pressure gradients is not a common feature of CMR, as these gradients cannot be accurately evaluated in routine CMR.

Methods

Twenty-five CHD patients who were planned for combined clinical CMR and diagnostic and/or interventional catheterization were enrolled in the trial. Prior to inclusion, a specific procedure for catheterization in CMR was defined, encompassing the assessment of pressure and pressure gradients in the heart and great vessels.

Results

By the use of an MR-conditional guidewire we successfully measured specific pressure and pressure gradients in up to 92% of cases with liquid-filled catheters which were guided exclusively under CMR guidance. There were no guidewire-related adverse events, and guidewire guidance and manipulation of catheters were successful.

Conclusions

Using a MR-conditional guidewire assists in easily reaching targets in the heart and great vessels and makes the catheter itself visible, so that invasive blood pressure assessment by CMR guidance with liquid-filled catheters can be improved.

Keywords

Cardiovascular magnetic resonance (CMR); congenital heart disease (CHD); cardiac catheterization; magnetic resonance; pressure; guidewire.

SUBMITTER: Meierhofer C 

PROVIDER: S-EPMC8748488 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

2017-01-17 | E-MTAB-5406 | biostudies-arrayexpress
| S-EPMC6934499 | biostudies-literature
| S-EPMC7048617 | biostudies-literature
| S-EPMC7653008 | biostudies-literature
| S-EPMC9636181 | biostudies-literature
| S-EPMC3219582 | biostudies-other
| S-EPMC10864253 | biostudies-literature
| S-EPMC9883634 | biostudies-literature
| S-EPMC2747120 | biostudies-other