Evaluation of acquisition modes for semi-quantitative analysis by targeted and untargeted mass spectrometry.


ABSTRACT:

Rationale

Analyte quantitation by mass spectrometry underpins a diverse range of scientific endeavors. The fast-growing field of mass spectrometer development has resulted in several targeted and untargeted acquisition modes suitable for these applications. By characterizing the acquisition methods available on an ion mobility (IM)-enabled orthogonal acceleration time-of-flight (oa-ToF) instrument, the optimum modes for analyte semi-quantitation can be deduced.

Methods

Serial dilutions of commercial metabolite, peptide, or cross-linked peptide analytes were prepared in matrices of human urine or Escherichia coli digest. Each analyte dilution was introduced into an IM separation-enabled oa-ToF mass spectrometer by reversed-phase liquid chromatography and electrospray ionization. Data were acquired for each sample in duplicate using nine different acquisition modes, including four IM-enabled acquisitions modes, available on the mass spectrometer.

Results

Five (metabolite) or seven (peptide/cross-linked peptide) point calibration curves were prepared for analytes across each of the acquisition modes. A nonlinear response was observed at high concentrations for some modes, attributed to saturation effects. Two correction methods, one MS1 isotope-correction and one MS2 ion intensity-correction, were applied to address this observation, resulting in an up to twofold increase in dynamic range. By averaging the semi-quantitative results across analyte classes, two parameters, linear dynamic range (LDR) and lower limit of quantification (LLOQ), were determined to evaluate each mode.

Conclusion

A comparison of the acquisition modes revealed that data-independent acquisition and parallel reaction monitoring methods are most robust for semi-quantitation when considering achievable LDR and LLOQ. IM-enabled modes exhibited sensitivity increases, but a simultaneous reduction in dynamic range required correction methods to recover. These findings will assist users in identifying the optimum acquisition mode for their analyte quantitation needs, supporting a diverse range of applications and providing guidance for future acquisition mode developments.

SUBMITTER: Britt HM 

PROVIDER: S-EPMC9287043 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7697060 | biostudies-literature
| S-EPMC7722469 | biostudies-literature
| S-EPMC6442367 | biostudies-literature
2005-09-20 | GSE2744 | GEO
| S-EPMC8583964 | biostudies-literature
| S-EPMC5924490 | biostudies-literature
| S-EPMC5623951 | biostudies-literature
| S-EPMC8336448 | biostudies-literature
| S-EPMC7465979 | biostudies-literature
2021-07-20 | GSE163326 | GEO