ABSTRACT: Background
Uncertainty regarding the reproducibility of the apparent diffusion coefficient (ADC) hampers the use of quantitative diffusion-weighted imaging (DWI) in evaluation of the prostate with magnetic resonance imaging MRI. The quantitative imaging biomarkers alliance (QIBA) profile for quantitative DWI claims a within-subject coefficient of variation (wCV) for prostate lesion ADC of 0.17. Improved understanding of ADC reproducibility would aid the use of quantitative diffusion in prostate MRI evaluation.Purpose
Evaluation of the repeatability (same-day) and reproducibility (multi-day) of whole-prostate and focal-lesion ADC assessment in a multi-site setting.Study type
Prospective multi-institutional.Subjects
Twenty-nine males, ages 53 to 80 (median 63) years, following diagnosis of prostate cancer, 10 with focal lesions.Field strength/sequence
3T, single-shot spin-echo diffusion-weighted echo-planar sequence with four b-values.Assessment
Sites qualified for the study using an ice-water phantom with known ADC. Readers performed DWI analyses at visit 1 ("V1") and visit 2 ("V2," 2-14 days after V1), where V2 comprised scans before ("V2pre") and after ("V2post") a "coffee-break" interval with subject removal and repositioning. A single reader segmented the whole prostate. Two readers separately placed region-of-interests for focal lesions.Statistical tests
Reproducibility and repeatability coefficients for whole prostate and focal lesions derived from median pixel ADC. We estimated the wCV and 95% confidence interval using a variance stabilizing transformation and assessed interreader reliability of focal lesion ADC using the intraclass correlation coefficient (ICC).Results
The ADC biases from b0 -b600 and b0 -b800 phantom scans averaged 1.32% and 1.44%, respectively; mean b-value dependence was 0.188%. Repeatability and reproducibility of whole prostate median pixel ADC both yielded wCVs of 0.033 (N = 29). In 10 subjects with an evaluable focal lesion, the individual reader wCVs were 0.148 and 0.074 (repeatability) and 0.137 and 0.078 (reproducibility). All time points demonstrated good to excellent interreader reliability for focal lesion ADC (ICCV1 = 0.89; ICCV2pre = 0.76; ICCV2post = 0.94).Data conclusion
This study met the QIBA claim for prostate ADC. Test-retest repeatability and multi-day reproducibility were largely equivalent. Interreader reliability for focal lesion ADC was high across time points.Level of evidence
1 TECHNICAL EFFICACY: Stage 2 TOC CATEGORY: Pelvis.
SUBMITTER: Boss MA
PROVIDER: S-EPMC9363527 | biostudies-literature | 2022 Sep
REPOSITORIES: biostudies-literature
Boss Michael A MA Snyder Bradley S BS Kim Eunhee E Flamini Dena D Englander Sarah S Sundaram Karthik M KM Gumpeni Naveen N Palmer Suzanne L SL Choi Haesun H Froemming Adam T AT Persigehl Thorsten T Davenport Matthew S MS Malyarenko Dariya D Chenevert Thomas L TL Rosen Mark A MA
Journal of magnetic resonance imaging : JMRI 20220210 3
<h4>Background</h4>Uncertainty regarding the reproducibility of the apparent diffusion coefficient (ADC) hampers the use of quantitative diffusion-weighted imaging (DWI) in evaluation of the prostate with magnetic resonance imaging MRI. The quantitative imaging biomarkers alliance (QIBA) profile for quantitative DWI claims a within-subject coefficient of variation (wCV) for prostate lesion ADC of 0.17. Improved understanding of ADC reproducibility would aid the use of quantitative diffusion in p ...[more]