Harmonizing platelet function analyzer testing and reporting in a large laboratory network.


ABSTRACT:

Introduction

The platelet function analyzer (PFA) is a popular platelet function screening instrument, highly sensitive to von Willebrand disease (VWD) and to aspirin therapy, with moderate sensitivity to defects in platelet function and/or deficiencies in platelet number. There are two models, the original PFA-100 and the contemporary PFA-200. Normal reference ranges (NRRs) provided by the manufacturer are the same for both models, instead being based on the type of test cartridge, for which there are two main ones: collagen/epinephrine (C/Epi) and collagen/adenosine diphosphate (C/ADP).

Methods

Comparative evaluations of PFA testing and reporting in six different sites of a large pathology network, aiming to harmonize NRRs and test reporting across all network sites. A separate comparative study of testing a range of samples (n > 150) on a PFA-100 versus that on a PFA-200. Review of contemporary literature.

Results

Each site was identified to have a different reporting NRR, which after consolidating data permitted establishment of an agreed harmonized NRR for use across the network (C/Epi: 90-160; C/ADP: 70-124; based on n > 180). Similarly, each site reported and interpreted results in different ways, and after discussion and consolidation, a harmonized approach to interpretation and reporting was achieved. The separate comparative study of PFA-100 versus PFA-200 testing confirmed instrument equivalence.

Conclusion

We achieved harmonized NRRs and reporting for PFA testing across a large pathology network. Our approach may be useful for other laboratory networks wishing to harmonize PFA testing.

SUBMITTER: Favaloro EJ 

PROVIDER: S-EPMC9545980 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Harmonizing platelet function analyzer testing and reporting in a large laboratory network.

Favaloro Emmanuel J EJ   Mohammed Soma S   Vong Ronny R   Chapman Kent K   Kershaw Geoffrey G   Just Sarah S   Connelly Lynne L   Ryan Michael M   Zebeljan Diane D   Brighton Timothy T   Pasalic Leonardo L  

International journal of laboratory hematology 20220626 5


<h4>Introduction</h4>The platelet function analyzer (PFA) is a popular platelet function screening instrument, highly sensitive to von Willebrand disease (VWD) and to aspirin therapy, with moderate sensitivity to defects in platelet function and/or deficiencies in platelet number. There are two models, the original PFA-100 and the contemporary PFA-200. Normal reference ranges (NRRs) provided by the manufacturer are the same for both models, instead being based on the type of test cartridge, for  ...[more]

Similar Datasets

| S-EPMC6715187 | biostudies-literature
| S-EPMC9578582 | biostudies-literature
| S-EPMC10621536 | biostudies-literature
| S-EPMC4975182 | biostudies-other
| S-EPMC8041198 | biostudies-literature
| S-EPMC7962106 | biostudies-literature
| S-EPMC9311435 | biostudies-literature
| S-EPMC3965432 | biostudies-literature
| S-EPMC9925488 | biostudies-literature
| S-EPMC9540416 | biostudies-literature