Pre-clinical validation of a turbine-based ventilator for invasive ventilation-The ACUTE-19 ventilator.


ABSTRACT:

Background

The Severe Acute Respiratory Syndrome (SARS)-Coronavirus 2 (CoV-2) pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for invasive ventilation.

Methods

Validation consisted of (a) testing tidal volume (VT) delivery in 11 simulated models, with various resistances and compliances; (b) comparison with a commercial ventilator (VIVO-50) adapting the United Kingdom Medicines and Healthcare products Regulatory Agency-recommendations for rapidly manufactured ventilators; and (c) in vivo testing in a sheep before and after inducing acute respiratory distress syndrome (ARDS) by saline lavage.

Results

Differences in VT in the simulated models were marginally different (largest difference 33ml [95%-confidence interval (CI) 31-36]; P<.001ml). Plateau pressure (Pplat) was not different (-0.3cmH2O [95%-CI -0.9 to 0.3]; P=.409), and positive end-expiratory pressure (PEEP) was marginally different (0.3 cmH2O [95%-CI 0.2 to 0.3]; P<.001) between the ACUTE-19 and the commercial ventilator. Bland-Altman analyses showed good agreement (mean bias, -0.29, [limits of agreement, 0.82 to -1.42], and mean bias 0.56 [limits of agreement, 1.94 to -0.81], at a Pplat of 15 and 30cmH2O, respectively). The ACUTE-19 achieved optimal oxygenation and ventilation before and after ARDS induction.

Conclusions

The ACUTE-19 performed accurately in simulated and animal models yielding a comparable performance with a VIVO-50 commercial device. The acute 19 can provide the basis for the development of a future affordable commercial ventilator.

SUBMITTER: Alonso-Inigo JM 

PROVIDER: S-EPMC9639442 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>The Severe Acute Respiratory Syndrome (SARS)-Coronavirus 2 (CoV-2) pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for invasive ventilation.<h4>Methods</h4>Validation consisted of (a) testing tidal volume (V<sub>T</sub>) delivery in 11 simulated models, with various resistances and compliances;  ...[more]

Similar Datasets

| S-EPMC9617684 | biostudies-literature
| S-EPMC9468466 | biostudies-literature
| S-EPMC11499821 | biostudies-literature
| S-EPMC2517084 | biostudies-literature
| S-EPMC7431581 | biostudies-literature
| S-EPMC7481711 | biostudies-literature
| S-EPMC5543010 | biostudies-other
| S-EPMC11525081 | biostudies-literature
| S-EPMC7605199 | biostudies-literature
| S-EPMC10464197 | biostudies-literature