ABSTRACT: Introduction
Indoxyl sulfate (IS), a protein-bound uremic toxin, is associated with kidney function and chronic kidney disease (CKD)-related complications. Currently, serum IS levels are primarily quantified using mass spectrometry-based methods, which are not feasible for routine clinical examinations.Methods
The efficiencies of three commercial ELISA kits in determination of serum IS were validated by comparing with ultra-performance liquid chromatography (UPLC)-MS/MS-based method using Bland-Altman analysis. The associations between kidney parameters and serum IS levels determined by ELISA kit from Leadgene and UPLC-MS/MS were evaluated by Spearman correlation coefficient in a CKD validation cohort.Results
ELISA kit from Leadgene showed clinical agreement with UPLC-MS/MS in the determination of serum IS levels (p = 0.084). In patients with CKD, Spearman's correlation analysis revealed a perfect correlation between the IS levels determined using the Leadgene ELISA kit and UPLC-MS/MS (r = 0.964, p < 0.0001). IS levels determined using the Leadgene ELISA kit were associated with the estimated glomerular filtration rate (r = -0.772, p < 0.0001) and serum creatinine concentration (r = 0.824, p < 0.0001) in patients with CKD, and on dialysis (r = 0.557, p = 0.006).Conclusions
The Leadgene ELISA kit exhibits comparable efficacy to UPLC-MS/MS in quantifying serum IS levels, supporting that ELISA would be a personalized method for monitoring the dynamic changes in serum IS levels in dialysis patients to prevent the progression of CKD.
SUBMITTER: Duan S
PROVIDER: S-EPMC9801083 | biostudies-literature | 2022 Dec
REPOSITORIES: biostudies-literature
Duan Shuangdi S Pi Jiayi J Wang Chun-Hsiang CH Hou Yi-Chou YC Andy Lee Chung-Ying CY Lin Cheng-Jui CJ Shi Liyang L Young Kung-Chia KC Sun Hung-Yu HY
Heliyon 20221214 12
<h4>Introduction</h4>Indoxyl sulfate (IS), a protein-bound uremic toxin, is associated with kidney function and chronic kidney disease (CKD)-related complications. Currently, serum IS levels are primarily quantified using mass spectrometry-based methods, which are not feasible for routine clinical examinations.<h4>Methods</h4>The efficiencies of three commercial ELISA kits in determination of serum IS were validated by comparing with ultra-performance liquid chromatography (UPLC)-MS/MS-based met ...[more]