Acceptability and Adherence to Home, Kiosk, and Clinic Blood Pressure Measurement Compared to 24-H Ambulatory Monitoring.


ABSTRACT:

Background

The US Preventive Services Task Force recommends measuring blood pressure (BP) outside of clinic/office settings. While various options are available, including home devices, BP kiosks, and 24-h ambulatory BP monitoring (ABPM), understanding patient acceptability and adherence is a critical factor for implementation.

Objective

To compare the acceptability and adherence of clinic, home, kiosk, and ABPM measurement.

Design

Comparative diagnostic accuracy study which randomized adults to one of three BP measurement arms: clinic, home, and kiosk. ABPM was conducted on all participants.

Participants

Adults (18-85 years) receiving care at 12 Kaiser Permanente Washington primary care clinics (Washington State, USA) with a high BP (≥ 138 mmHg systolic or ≥ 88 mmHg diastolic) in the electronic health record with no hypertension diagnosis and on no hypertensive medications and with high BP at a research screening visit.

Measures

Patient acceptability was measured using a validated survey which was used to calculate an overall acceptability score (range 1-7) at baseline, after completing their assigned BP measurement intervention, and after completing ABPM. Adherence was defined based on the pre-specified number of BP measurements completed.

Key results

Five hundred ten participants were randomized (mean age 59 years), with mean BP of 150/88. Overall acceptability score was highest (i.e. most acceptable) for Home BP (mean 6.2, SD 0.7) and lowest (least acceptable) for ABPM (mean 5.0, SD 1.0); scores were intermediate for Clinic (5.5, SD 1.1) and Kiosk (5.4, SD 1.0). Adherence was higher for Home (154/170, 90.6%) and Clinic (150/172, 87.2%) than for Kiosk (114/168, 67.9%)). The majority of participants (467/510, 91.6%) were adherent to ABPM.

Conclusions

Participants found home BP measurement most acceptable followed by clinic, BP kiosks, and ABPM. Our findings, coupled with recent evidence regarding the accuracy of home BP measurement, further support the routine use of home-based BP measurement in primary care practice in the US.

Trial registration

ClinicalTrials.gov NCT03130257 https://clinicaltrials.gov/ct2/show/NCT03130257.

SUBMITTER: Thompson MJ 

PROVIDER: S-EPMC9845022 | biostudies-literature | 2023 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Acceptability and Adherence to Home, Kiosk, and Clinic Blood Pressure Measurement Compared to 24-H Ambulatory Monitoring.

Thompson Matthew J MJ   Anderson Melissa L ML   Cook Andrea J AJ   Ehrlich Kelly K   Hall Yoshio N YN   Hsu Clarissa C   Margolis Karen L KL   McClure Jennifer B JB   Munson Sean A SA   Green Beverly B BB  

Journal of general internal medicine 20230117 8


<h4>Background</h4>The US Preventive Services Task Force recommends measuring blood pressure (BP) outside of clinic/office settings. While various options are available, including home devices, BP kiosks, and 24-h ambulatory BP monitoring (ABPM), understanding patient acceptability and adherence is a critical factor for implementation.<h4>Objective</h4>To compare the acceptability and adherence of clinic, home, kiosk, and ABPM measurement.<h4>Design</h4>Comparative diagnostic accuracy study whic  ...[more]

Similar Datasets

| S-EPMC7067555 | biostudies-literature
| S-EPMC9929017 | biostudies-literature
| S-EPMC3122300 | biostudies-literature
| S-EPMC10864248 | biostudies-literature
| S-EPMC9485334 | biostudies-literature
| S-EPMC10765747 | biostudies-literature
| S-EPMC7560573 | biostudies-literature
| S-EPMC9444260 | biostudies-literature
| S-EPMC6207212 | biostudies-literature
| S-EPMC3915515 | biostudies-literature