A broadband imaging system for research applications.
Ontology highlight
ABSTRACT: We have developed a compact, computer-piloted, high sensitivity broadband imaging system for laboratory research that is compatible with various detectors. Mirror optics allow application from the visible to the far infrared spectral range. A prototype tested in conjunction with a mercury cadmium telluride detector exhibits a peak detectivity of 6.7x10(10) cm Hz(1/2)/W at a wavelength of 11.8 microm. Temperature and spatial resolutions of 0.06 K and 1.6 mrad, respectively, were demonstrated.