Inkjet metrology: high-accuracy mass measurements of microdroplets produced by a drop-on-demand dispenser.
Ontology highlight
ABSTRACT: We describe gravimetric methods for measuring the mass of droplets generated by a drop-on-demand (DOD) microdispenser. Droplets are deposited, either continuously at a known frequency or as a burst of known number, into a cylinder positioned on a submicrogram balance. Mass measurements are acquired precisely by computer, and results are corrected for evaporation. Capabilities are demonstrated using isobutyl alcohol droplets. For ejection rates greater than 100 Hz, the repeatability of droplet mass measurements was 0.2%, while the combined relative standard uncertainty (u(c)) was 0.9%. When bursts of droplets were dispensed, the limit of quantitation was 72 microg (1490 droplets) with u(c) = 1.0%. Individual droplet size in a burst was evaluated by high-speed videography. Diameters were consistent from the tenth droplet onward, and the mass of an individual droplet was best estimated by the average droplet mass with a combined uncertainty of about 1%. Diameters of the first several droplets were anomalous, but their contribution was accounted for when dispensing bursts. Above the limits of quantitation, the gravimetric methods provided statistically equivalent results and permit detailed study of operational factors that influence droplet mass during dispensing, including the development of reliable microassays and standard materials using DOD technologies.
SUBMITTER: Verkouteren RM
PROVIDER: S-EPMC2762308 | biostudies-other | 2009 Oct
REPOSITORIES: biostudies-other
ACCESS DATA