Unknown

Dataset Information

0

Improved R2* measurement accuracy with absolute SNR truncation and optimal coil combination.


ABSTRACT: Accurate R2* measurements are critical for many abdominal imaging applications. Conventionally, R2* maps are derived via the monoexponential fitting of signal decay within a series of gradient-echo (GRE) images reconstructed from multichannel datasets combined using a root sum-of-squares (RSS) approach. However, the noise bias at low-SNR TEs from RSS-reconstructed data often causes the underestimation of R2* values. In phantom, ex vivo animal model and normal volunteer studies, we investigated the accuracy of low-SNR R2* measurement when combining truncation and coil combination methods. The accuracy for R2* estimations was shown to be affected by the intrinsic R2* value, SNR level and the chosen reconstruction method. The R2* estimation error was found to decrease with increasing SNR level, decreasing R2* value and the use of the optimal B1-weighted combined (OBC) image reconstruction method. Data truncation based on rigorous voxel-wise SNR estimates can reduce R2* measurement error in the setting of low SNR with fast signal decay. When optimal SNR truncation thresholds are unknown, the OBC method can provide optimal R2* measurements given the minimal truncation requirements.

SUBMITTER: Yin X 

PROVIDER: S-EPMC3043554 | biostudies-other | 2010 Dec

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC8575292 | biostudies-literature
2009-10-08 | GSE13744 | GEO
2010-04-08 | E-GEOD-13744 | biostudies-arrayexpress
| S-EPMC4256073 | biostudies-literature
| S-EPMC5908153 | biostudies-literature
2023-01-01 | GSE189142 | GEO
| S-EPMC4529141 | biostudies-literature
| S-EPMC8466921 | biostudies-literature
| S-EPMC4729197 | biostudies-literature
| S-EPMC7806663 | biostudies-literature