ABSTRACT: The present study aimed to examine the effect of electrode configuration, specifically monopolar (MP) or bipolar (BP) stimulation, on place pitch discrimination in cochlear implants (CIs). Twelve subjects implanted with the Nucleus Freedom device were presented with various pairs of stimulation across the electrode array, with varying degrees of distance between stimulation sites, and asked to judge the higher of the two in pitch. Each pair was presented either in the same mode or in different modes of stimulation for the within-mode or across-mode condition, respectively, at least 20 times. The result of the within-mode condition revealed that subjects, on average, were able to discriminate pitches significantly better in MP than in BP, with the sensitivity index (d') for adjacent channels of 1.2 for MP and 0.8 for BP. The result of the across-mode condition revealed that while individual variability existed, there was a strong tendency for CI subjects to perceive a higher pitch in BP stimulation than in MP for a similar site of stimulation. In other words, an MP channel needed to be shifted in a basal direction by as much as two electrodes on average to elicit a pitch comparable to that of a BP channel.
SUBMITTER: Kwon BJ
PROVIDER: S-EPMC3135145 | biostudies-other | 2011 Jun
REPOSITORIES: biostudies-other