Unknown

Dataset Information

0

Retinal oximeter for the blue-green oximetry technique.


ABSTRACT: Retinal oximetry offers potential for noninvasive assessment of central venous oxyhemoglobin saturation (SO(2)) via the retinal vessels but requires a calibrated accuracy of ±3% saturation in order to be clinically useful. Prior oximeter designs have been hampered by poor saturation calibration accuracy. We demonstrate that the blue-green oximetry (BGO) technique can provide accuracy within ±3% in swine when multiply scattered light from blood within a retinal vessel is isolated. A noninvasive on-axis scanning retinal oximeter (ROx-3) is constructed that generates a multiwavelength image in the range required for BGO. A field stop in the detection pathway is used in conjunction with an anticonfocal bisecting wire to remove specular vessel reflections and isolate multiply backscattered light from the blood column within a retinal vessel. This design is tested on an enucleated swine eye vessel and a retinal vein in a human volunteer with retinal SO(2) measurements of ∼1 and ∼65%, respectively. These saturations, calculated using the calibration line from earlier work, are internally consistent with a standard error of the mean of ±2% SO(2). The absolute measures are well within the expected saturation range for the site (-1 and 63%). This is the first demonstration of noninvasive on-axis BGO retinal oximetry.

SUBMITTER: Denninghoff KR 

PROVIDER: S-EPMC3206928 | biostudies-other | 2011 Oct

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC7233590 | biostudies-literature
| S-EPMC8675673 | biostudies-literature
| S-EPMC4951009 | biostudies-literature
| S-EPMC4740427 | biostudies-literature
2021-11-20 | GSE189122 | GEO
2024-09-30 | GSE271876 | GEO
| S-EPMC7260968 | biostudies-literature
| S-EPMC9716545 | biostudies-literature
| S-EPMC6462473 | biostudies-literature
| S-EPMC6033873 | biostudies-literature