Unknown

Dataset Information

0

Rapid multi sample DNA amplification using rotary-linear polymerase chain reaction device (PCRDisc).


ABSTRACT: Multiple sample DNA amplification was done by using a novel rotary-linear motion polymerase chain reaction (PCR) device. A simple compact disc was used to create the stationary sample chambers which are individually temperature controlled. The PCR was performed by shuttling the samples to different temperature zones by using a combined rotary-linear movement of the disc. The device was successfully used to amplify up to 12 samples in less than 30 min with a sample volume of 5 μl. A simple spring loaded heater mechanism was introduced to enable good thermal contact between the samples and the heaters. Each of the heater temperatures are controlled by using a simple proportional-integral-derivative pulse width modulation control system. The results show a good improvement in the amplification rate and duration of the samples. The reagent volume used was reduced to nearly 25% of that used in conventional method.

SUBMITTER: Sugumar D 

PROVIDER: S-EPMC3370399 | biostudies-other | 2012 Mar

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC297692 | biostudies-other
| S-EPMC4337463 | biostudies-literature
| S-EPMC3555095 | biostudies-literature
| S-EPMC3634412 | biostudies-literature
| S-EPMC8491538 | biostudies-literature
| S-EPMC5161023 | biostudies-literature
| S-EPMC6764677 | biostudies-literature
| S-EPMC7112770 | biostudies-literature
| S-EPMC5885343 | biostudies-literature
| S-EPMC7161830 | biostudies-literature