Fabrication and characterization of medical grade polyurethane composite catheters for near-infrared imaging.


ABSTRACT: Peripherally inserted central catheters (PICCs) are hollow polymeric tubes that transport nutrients, blood and medications to neonates. To determine proper PICC placement, frequent X-ray imaging of neonates is performed. Because X-rays pose severe health risks to neonates, safer alternatives are needed. We hypothesize that near infrared (NIR) polymer composites can be fabricated into catheters by incorporating a fluorescent dye (IRDye 800CW) and visualized using NIR imaging. To fabricate catheters, polymer and dye are dry mixed and pressed, sectioned, and extruded to produce hollow tubes. We analyzed surface roughness, stiffness, dye retention, NIR contrast intensity, and biocompatibility. The extrusion process did not significantly alter the mechanical properties of the polymer composites. Over a period of 23 days, only 6.35 ± 5.08% dye leached out of catheters. The addition of 0.025 wt% dye resulted in a 14-fold contrast enhancement producing clear PICC images at 1 cm under a tissue equivalent. The addition of IRDye 800CW did not alter the biocompatibility of the polymer and did not increase adhesion of cells to the surface. We successfully demonstrated that catheters can be imaged without the use of harmful radiation and still maintain the same properties as the unaltered medical grade equivalent.

SUBMITTER: Stevenson AT 

PROVIDER: S-EPMC4417621 | biostudies-other | 2015 Jun

REPOSITORIES: biostudies-other

altmetric image

Publications

Fabrication and characterization of medical grade polyurethane composite catheters for near-infrared imaging.

Stevenson André T AT   Reese Laura M LM   Hill Tanner K TK   McGuire Jeffrey J   Mohs Aaron M AM   Shekhar Raj R   Bickford Lissett R LR   Whittington Abby R AR  

Biomaterials 20150407


Peripherally inserted central catheters (PICCs) are hollow polymeric tubes that transport nutrients, blood and medications to neonates. To determine proper PICC placement, frequent X-ray imaging of neonates is performed. Because X-rays pose severe health risks to neonates, safer alternatives are needed. We hypothesize that near infrared (NIR) polymer composites can be fabricated into catheters by incorporating a fluorescent dye (IRDye 800CW) and visualized using NIR imaging. To fabricate cathete  ...[more]

Similar Datasets

| S-EPMC4312223 | biostudies-literature
| S-EPMC8339317 | biostudies-literature
| S-EPMC8875655 | biostudies-literature
| S-EPMC5061038 | biostudies-literature
| S-EPMC6401970 | biostudies-literature
| S-EPMC8450770 | biostudies-literature
| S-EPMC8686222 | biostudies-literature
| S-EPMC5744357 | biostudies-literature
| S-EPMC2646119 | biostudies-literature
| S-EPMC4575506 | biostudies-literature