Coronary CT Angiography: Variability of CT Scanners and Readers in Measurement of Plaque Volume.


ABSTRACT: Purpose To determine reader and computed tomography (CT) scan variability for measurement of coronary plaque volume. Materials and Methods This HIPAA-compliant study followed Standards for Reporting of Diagnostic Accuracy guidelines. Baseline coronary CT angiography was performed in 40 prospectively enrolled subjects (mean age, 67 years ± 6 [standard deviation]) with asymptomatic hyperlipidemia by using a 320-detector row scanner (Aquilion One Vision; Toshiba, Otawara, Japan). Twenty of these subjects underwent coronary CT angiography repeated on a separate day with the same CT scanner (Toshiba, group 1); 20 subjects underwent repeat CT performed with a different CT scanner (Somatom Force; Siemens, Forchheim, Germany [group 2]). Intraclass correlation coefficients (ICCs) and Bland-Altman analysis were used to assess interreader, intrareader, and interstudy reproducibility. Results Baseline and repeat coronary CT angiography scans were acquired within 19 days ± 6. Interreader and intrareader agreement rates were high for total, calcified, and noncalcified plaques for both CT scanners (all ICCs ? 0.96) without bias. Scanner variability was ±18.4% (coefficient of variation) with same-vendor follow-up. However, scanner variability increased to ±29.9% with different-vendor follow-up. The sample size to detect a 5% change in noncalcified plaque volume with 90% power and an ? error of .05 was 286 subjects for same-CT scanner follow-up and 753 subjects with different-vendor follow-up. Conclusion State-of-the-art coronary CT angiography with same-vendor follow-up has good scan-rescan reproducibility, suggesting a role of coronary CT angiography in monitoring coronary artery plaque response to therapy. Differences between coronary CT angiography vendors resulted in lower scan-rescan reproducibility. © RSNA, 2016 Online supplemental material is available for this article.

SUBMITTER: Symons R 

PROVIDER: S-EPMC5131836 | biostudies-other | 2016 Dec

REPOSITORIES: biostudies-other

altmetric image

Publications

Coronary CT Angiography: Variability of CT Scanners and Readers in Measurement of Plaque Volume.

Symons Rolf R   Morris Justin Z JZ   Wu Colin O CO   Pourmorteza Amir A   Ahlman Mark A MA   Lima João A C JA   Chen Marcus Y MY   Mallek Marissa M   Sandfort Veit V   Bluemke David A DA  

Radiology 20160916 3


Purpose To determine reader and computed tomography (CT) scan variability for measurement of coronary plaque volume. Materials and Methods This HIPAA-compliant study followed Standards for Reporting of Diagnostic Accuracy guidelines. Baseline coronary CT angiography was performed in 40 prospectively enrolled subjects (mean age, 67 years ± 6 [standard deviation]) with asymptomatic hyperlipidemia by using a 320-detector row scanner (Aquilion One Vision; Toshiba, Otawara, Japan). Twenty of these su  ...[more]

Similar Datasets

2022-06-29 | GSE202625 | GEO
| S-EPMC4613877 | biostudies-literature
| S-EPMC6328572 | biostudies-literature
| S-EPMC7089811 | biostudies-literature
| S-EPMC4858195 | biostudies-literature
| S-EPMC8382040 | biostudies-literature
| S-EPMC4455680 | biostudies-literature
| S-EPMC6939749 | biostudies-literature
| S-EPMC4556695 | biostudies-literature
| S-EPMC7977955 | biostudies-literature