Investigating the effect of a magnetic field on dose distributions at phantom-air interfaces using PRESAGE® 3D dosimeter and Monte Carlo simulations.


ABSTRACT: Dosimetric quality assurance (QA) of the new Elekta Unity (MR-linac) will differ from the QA performed of a conventional linac due to the constant magnetic field, which creates an electron return effect (ERE). In this work we aim to validate PRESAGE® dosimetry in a transverse magnetic field, and assess its use to validate the research version of the Monaco TPS of the MR-linac. Cylindrical samples of PRESAGE® 3D dosimeter separated by an air gap were irradiated with a cobalt-60 unit, while placed between the poles of an electromagnet at 0.5 T and 1.5 T. This set-up was simulated in EGSnrc/Cavity Monte Carlo (MC) code and relative dose distributions were compared with measurements using 1D and 2D gamma criteria of 3% and 1.5 mm. The irradiation conditions were adapted for the MR-linac and compared with Monaco TPS simulations. Measured and EGSnrc/Cavity simulated profiles showed good agreement with a gamma passing rate of 99.9% for 0.5 T and 99.8% for 1.5 T. Measurements on the MR-linac also compared well with Monaco TPS simulations, with a gamma passing rate of 98.4% at 1.5 T. Results demonstrated that PRESAGE® can accurately measure dose and detect the ERE, encouraging its use as a QA tool to validate the Monaco TPS of the MR-linac for clinically relevant dose distributions at tissue-air boundaries.

SUBMITTER: Costa F 

PROVIDER: S-EPMC5964337 | biostudies-other | 2018 Feb

REPOSITORIES: biostudies-other

altmetric image

Publications

Investigating the effect of a magnetic field on dose distributions at phantom-air interfaces using PRESAGE<sup>®</sup> 3D dosimeter and Monte Carlo simulations.

Costa Filipa F   Doran Simon J SJ   Hanson Ian M IM   Nill Simeon S   Billas Ilias I   Shipley David D   Duane Simon S   Adamovics John J   Oelfke Uwe U  

Physics in medicine and biology 20180226 5


Dosimetric quality assurance (QA) of the new Elekta Unity (MR-linac) will differ from the QA performed of a conventional linac due to the constant magnetic field, which creates an electron return effect (ERE). In this work we aim to validate PRESAGE<sup>®</sup> dosimetry in a transverse magnetic field, and assess its use to validate the research version of the Monaco TPS of the MR-linac. Cylindrical samples of PRESAGE<sup>®</sup> 3D dosimeter separated by an air gap were irradiated with a cobalt  ...[more]

Similar Datasets

| S-EPMC5711240 | biostudies-literature
| S-EPMC2999362 | biostudies-literature
| S-EPMC6465329 | biostudies-literature
| S-EPMC9263979 | biostudies-literature
| S-EPMC5875513 | biostudies-literature
| S-EPMC5553226 | biostudies-other
| S-EPMC6432162 | biostudies-literature
| S-EPMC2998598 | biostudies-literature
| S-EPMC4919768 | biostudies-literature