Comparison of unit resolution versus high-resolution accurate mass for parallel reaction monitoring


ABSTRACT: Parallel reaction monitoring (PRM) is an increasingly popular alternative to selected reaction monitoring (SRM) for targeted proteomics. PRM’s strengths over SRM are that it monitors all product ions in a single spectrum, thus eliminating the need to select interference-free product ions prior to data acquisition, and that it is most frequently performed on high-resolution instruments, such as quadrupole-orbitrap and quadrupole-time of flight instruments. Here, we show that the primary advantage of PRM is the ability to monitor all transitions in parallel, and that high-resolution data are not necessary to obtain high quality quantitative data. We run the same scheduled PRM assay, measuring 432 peptides from 126 plasma proteins, multiple times on a Orbitrap Eclipse Tribrid mass spectrometer, alternating separate liquid chromatography-tandem mass spectrometry runs between the high resolution Orbitrap and the unit resolution linear ion trap for PRM. We find that both mass analyzers have similar technical precision, and that the linear ion trap’s superior sensitivity gives it better lower limits of quantitation on over 62% of peptides in the assay.

ORGANISM(S): Homo Sapiens

SUBMITTER: Lilian Heil  

PROVIDER: PXD023334 | panorama | Thu Aug 19 00:00:00 BST 2021

REPOSITORIES: PanoramaPublic

altmetric image

Publications

Comparison of Unit Resolution Versus High-Resolution Accurate Mass for Parallel Reaction Monitoring.

Heil Lilian R LR   Remes Philip M PM   MacCoss Michael J MJ  

Journal of proteome research 20210728 9


Parallel reaction monitoring (PRM) is an increasingly popular alternative to selected reaction monitoring (SRM) for targeted proteomics. PRM's strengths over SRM are that it monitors all product ions in a single spectrum, thus eliminating the need to select interference-free product ions prior to data acquisition, and that it is most frequently performed on high-resolution instruments, such as quadrupole-orbitrap and quadrupole-time-of-flight instruments. Here, we show that the primary advantage  ...[more]

Similar Datasets

2014-06-10 | PXD001047 | Pride
2018-07-05 | PXD007601 | Pride
2018-07-05 | PXD006615 | Pride
2022-03-10 | PXD006569 | Pride
2018-07-05 | PXD006618 | Pride
2017-08-24 | PXD006236 | Pride
2018-10-05 | PXD011009 | Pride
2024-03-29 | PXD049028 | Pride
2022-04-08 | PXD031143 | Pride
2020-06-19 | PXD017778 | Pride