ABSTRACT: Objectives
To assess the accuracy, repeatability, and reproducibility of T1 and T2 relaxation time measurements by three-dimensional magnetic resonance fingerprinting (3D MRF) using various dictionary resolutions.Methods
The ISMRM/NIST phantom was scanned daily for 10 days in two 3 T MR scanners using a 3D MRF sequence reconstructed using four dictionaries with varying step sizes and one dictionary with wider ranges. Thirty-nine healthy volunteers were enrolled: 20 subjects underwent whole-brain MRF scans in both scanners and the rest in one scanner. ROI/VOI analyses were performed on phantom and brain MRF maps. Accuracy, repeatability, and reproducibility metrics were calculated.Results
In the phantom study, all dictionaries showed high T1 linearity to the reference values (R2 > 0.99), repeatability (CV < 3%), and reproducibility (CV < 3%) with lower linearity (R2 > 0.98), repeatability (CV < 6%), and reproducibility (CV ≤ 4%) for T2 measurement. The volunteer study demonstrated high T1 reproducibility of within-subject CV (wCV) < 4% by all dictionaries with the same ranges, both in the brain parenchyma and CSF. Yet, reproducibility was moderate for T2 measurement (wCV < 8%). In CSF measurement, dictionaries with a smaller range showed a seemingly better reproducibility (T1, wCV 3%; T2, wCV 8%) than the much wider range dictionary (T1, wCV 5%; T2, wCV 13%). Truncated CSF relaxometry values were evident in smaller range dictionaries.Conclusions
The accuracy, repeatability, and reproducibility of 3D MRF across various dictionary resolutions were high for T1 and moderate for T2 measurements. A lower-resolution dictionary with a well-defined range may be adequate, thus significantly reducing the computational load.Key points
• A lower-resolution dictionary with a well-defined range may be sufficient for 3D MRF reconstruction. • CSF relaxation times might be underestimated due to truncation by the upper dictionary range. • Dictionary with a higher upper range might be advisable, especially for CSF evaluation and elderly subjects whose perivascular spaces are more prominent.
SUBMITTER: Wicaksono KP
PROVIDER: S-EPMC10017611 | biostudies-literature | 2023 Apr
REPOSITORIES: biostudies-literature
Wicaksono Krishna Pandu KP Fushimi Yasutaka Y Nakajima Satoshi S Sakata Akihiko A Okuchi Sachi S Hinoda Takuya T Oshima Sonoko S Otani Sayo S Tagawa Hiroshi H Urushibata Yuta Y Nakamoto Yuji Y
European radiology 20221124 4
<h4>Objectives</h4>To assess the accuracy, repeatability, and reproducibility of T<sub>1</sub> and T<sub>2</sub> relaxation time measurements by three-dimensional magnetic resonance fingerprinting (3D MRF) using various dictionary resolutions.<h4>Methods</h4>The ISMRM/NIST phantom was scanned daily for 10 days in two 3 T MR scanners using a 3D MRF sequence reconstructed using four dictionaries with varying step sizes and one dictionary with wider ranges. Thirty-nine healthy volunteers were enrol ...[more]