Unknown

Dataset Information

0

Development of a shutterless continuous rotation method using an X-ray CMOS detector for protein crystallography.


ABSTRACT: A new shutterless continuous rotation method using an X-ray complementary metal-oxide semiconductor (CMOS) detector has been developed for high-speed, precise data collection in protein crystallography. The principle of operation and the basic performance of the X-ray CMOS detector (Hamamatsu Photonics KK C10158DK) have been shown to be appropriate to the shutterless continuous rotation method. The data quality of the continuous rotation method is comparable to that of the conventional oscillation method using a CCD detector and, furthermore, the combination with fine ? slicing improves the data accuracy without increasing the data-collection time. The new method is more sensitive to diffraction intensity because of the narrow dynamic range of the CMOS detector. However, the strong diffraction spots were found to be precisely measured by recording them on successive multiple images by selecting an adequate rotation step. The new method has been used to successfully determine three protein structures by multi- and single-wavelength anomalous diffraction phasing and has thereby been proved applicable in protein crystallography. The apparatus and method may become a powerful tool at synchrotron protein crystallography beamlines with important potential across a wide range of X-ray wavelengths.

SUBMITTER: Hasegawa K 

PROVIDER: S-EPMC3246825 | biostudies-literature | 2009 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development of a shutterless continuous rotation method using an X-ray CMOS detector for protein crystallography.

Hasegawa Kazuya K   Hirata Kunio K   Shimizu Tetsuya T   Shimizu Nobutaka N   Hikima Takaaki T   Baba Seiki S   Kumasaka Takashi T   Yamamoto Masaki M  

Journal of applied crystallography 20091117 Pt 6


A new shutterless continuous rotation method using an X-ray complementary metal-oxide semiconductor (CMOS) detector has been developed for high-speed, precise data collection in protein crystallography. The principle of operation and the basic performance of the X-ray CMOS detector (Hamamatsu Photonics KK C10158DK) have been shown to be appropriate to the shutterless continuous rotation method. The data quality of the continuous rotation method is comparable to that of the conventional oscillati  ...[more]

Similar Datasets

| S-EPMC7064105 | biostudies-literature
| S-EPMC3182744 | biostudies-literature
| S-EPMC9850494 | biostudies-literature
| S-EPMC7201292 | biostudies-literature
| S-EPMC4052869 | biostudies-literature
| S-EPMC5947724 | biostudies-literature
| S-EPMC5013597 | biostudies-literature
| S-EPMC7044001 | biostudies-literature
| S-EPMC11100555 | biostudies-literature
| S-EPMC3786679 | biostudies-literature