Unknown

Dataset Information

0

Characterization and evaluation of the artemis camera for fluorescence-guided cancer surgery.


ABSTRACT: PURPOSE:Near-infrared (NIR) fluorescence imaging can provide the surgeon with real-time visualization of, e.g., tumor margins and lymph nodes. We describe and evaluate the Artemis, a novel, handheld NIR fluorescence camera. PROCEDURES:We evaluated minimal detectable cell numbers (FaDu-luc2, 7D12-IRDye 800CW), preclinical intraoperative detection of sentinel lymph nodes (SLN) using indocyanine green (ICG), and of orthotopic tongue tumors using 7D12-800CW. Results were compared with the Pearl imager. Clinically, three patients with liver metastases were imaged using ICG. RESULTS:Minimum detectable cell counts for Artemis and Pearl were 2 × 10(5) and 4 × 10(4) cells, respectively. In vivo, seven SLNs were detected in four mice with both cameras. Orthotopic OSC-19-luc2-cGFP tongue tumors were clearly identifiable, and a minimum FaDu-luc2 tumor size of 1 mm(3) could be identified. Six human malignant lesions were identified during three liver surgery procedures. CONCLUSIONS:Based on this study, the Artemis system has demonstrated its utility in fluorescence-guided cancer surgery.

SUBMITTER: van Driel PB 

PROVIDER: S-EPMC4422838 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC3409706 | biostudies-other
| S-EPMC3755013 | biostudies-other
| S-EPMC5572645 | biostudies-other
| S-EPMC6280663 | biostudies-literature
| S-EPMC7240319 | biostudies-literature
| S-EPMC4510011 | biostudies-literature
| S-EPMC6581234 | biostudies-literature
| S-EPMC8205604 | biostudies-literature
| S-EPMC4874040 | biostudies-literature
| S-EPMC5501082 | biostudies-literature