ABSTRACT: A novel electrochemically controlled release method for nitric oxide (NO) (based on electrochemical reduction of nitrite ions) is combined with an amperometric oxygen sensor within a dual lumen catheter configuration for the continuous in vivo sensing of the partial pressure of oxygen (PO2) in blood. The on-demand electrochemical NO generation/release method is shown to be fully compatible with amperometric PO2 sensing. The performance of the sensors is evaluated in rabbit veins and pig arteries for 7 and 21 h, respectively. Overall, the NO releasing sensors measure both venous and arterial PO2 values more accurately with an average deviation of -2 ± 11% and good correlation (R(2) = 0.97) with in vitro blood measurements, whereas the corresponding control sensors without NO release show an average deviation of -31 ± 28% and poor correlation (R(2) = 0.43) at time points >4 h after implantation in veins and >6 h in arteries. The NO releasing sensors induce less thrombus formation on the catheter surface in both veins and arteries (p < 0.05). This electrochemical NO generation/release method could offer a new and attractive means to improve the biocompatibility and performance of implantable chemical sensors.

SUBMITTER: Ren H 

PROVIDER: S-EPMC4556045 | biostudies-literature | 2015 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improved in vivo performance of amperometric oxygen (PO2) sensing catheters via electrochemical nitric oxide generation/release.

Ren Hang H   Coughlin Megan A MA   Major Terry C TC   Aiello Salvatore S   Rojas Pena Alvaro A   Bartlett Robert H RH   Meyerhoff Mark E ME  

Analytical chemistry 20150805 16


A novel electrochemically controlled release method for nitric oxide (NO) (based on electrochemical reduction of nitrite ions) is combined with an amperometric oxygen sensor within a dual lumen catheter configuration for the continuous in vivo sensing of the partial pressure of oxygen (PO2) in blood. The on-demand electrochemical NO generation/release method is shown to be fully compatible with amperometric PO2 sensing. The performance of the sensors is evaluated in rabbit veins and pig arteries  ...[more]

Similar Datasets

| S-EPMC2772994 | biostudies-literature
| S-EPMC6366661 | biostudies-literature
| S-EPMC5030524 | biostudies-literature
| S-EPMC8608891 | biostudies-literature
| S-EPMC7593647 | biostudies-literature
| S-EPMC2440544 | biostudies-literature
| S-EPMC4555983 | biostudies-literature
| S-EPMC2859168 | biostudies-literature
| S-EPMC4682557 | biostudies-literature
| S-EPMC7164781 | biostudies-literature