Reduced field of view imaging using a static second-order gradient for functional MRI applications.


ABSTRACT:

Purpose

Imaging using reduced FOV excitation allows higher resolution or signal-to-noise ratio (SNR) per scan time but often requires long radiofrequency pulses. The goal of this study was to improve a recent reduced field of view (FOV) method that uses a second-order shim gradient to decrease pulse length and evaluate its use in functional MRI (fMRI) applications.

Theory and methods

The method, which was initially limited to excite thin disc-shaped regions at the isocenter, was extended to excite thicker regions off the isocenter and produced accurate excitation profiles on a grid phantom. Visual stimulation fMRI scans were performed with full and reduced FOV. The resolution of the time series images and functional activation maps were assessed using the full-width half-maxima of the autocorrelation functions (FACFs) of the noise images and the activation map values, respectively.

Results

The resolution was higher in the reduced FOV time series images (4.1% ± 3.7% FACF reduction, P < 0.02) and functional activation maps (3.1% ± 3.4% FACF reduction, P < 0.01), but the SNR was lower (by 26.5% ± 16.9%). However, for a few subjects, the targeted region could not be localized to the reduced FOV due to the low Z2 gradient strength.

Conclusion

The results of this study suggest that the proposed method is feasible, though it would benefit from a stronger gradient coil.

SUBMITTER: Islam H 

PROVIDER: S-EPMC4583326 | biostudies-literature | 2016 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reduced field of view imaging using a static second-order gradient for functional MRI applications.

Islam Haisam H   Glover Gary H GH  

Magnetic resonance in medicine 20150325 2


<h4>Purpose</h4>Imaging using reduced FOV excitation allows higher resolution or signal-to-noise ratio (SNR) per scan time but often requires long radiofrequency pulses. The goal of this study was to improve a recent reduced field of view (FOV) method that uses a second-order shim gradient to decrease pulse length and evaluate its use in functional MRI (fMRI) applications.<h4>Theory and methods</h4>The method, which was initially limited to excite thin disc-shaped regions at the isocenter, was e  ...[more]

Similar Datasets

| S-EPMC5607067 | biostudies-literature
| S-EPMC4090294 | biostudies-literature
| S-EPMC6704114 | biostudies-literature
| S-EPMC3345893 | biostudies-literature
| S-EPMC4230370 | biostudies-literature
| S-EPMC5847324 | biostudies-literature
| S-EPMC4291562 | biostudies-literature
| S-EPMC7534367 | biostudies-literature