Diagnostic Accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care Hemoglobin Meters in a Central Laboratory and a Community Based Clinic in Durban, South Africa.


ABSTRACT: In South Africa, various point-of-care hemoglobin meters are used. However, the regulatory framework for approval, implementation and oversight of use of point-of-care hemoglobin meters is suboptimal. We assessed the diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 point-of-care hemoglobin meters, compared to a central laboratory based reference assay, in a central laboratory and a community based clinic in Durban, South Africa. Differences in performance of the point-of-care assays, compared to the reference assay, were more pronounced in the community based clinic. Results were reasonable for the HemoCue Hb 301, but poor for the STAT-Site MHgb and the URIT-12. Poor test performance of point-of-care hemoglobin meters, and inadequate evaluations and oversight in South Africa, leads to suboptimal clinical care and clinical research, and increased costs. There is a need for proper evaluation and quality assurance of point-of-care tests, the results of which should be made widely available to key stakeholders.

SUBMITTER: Jaggernath M 

PROVIDER: S-EPMC4821624 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic Accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care Hemoglobin Meters in a Central Laboratory and a Community Based Clinic in Durban, South Africa.

Jaggernath Manjeetha M   Naicker Rumallen R   Madurai Savathree S   Brockman Mark A MA   Ndung'u Thumbi T   Gelderblom Huub C HC  

PloS one 20160405 4


In South Africa, various point-of-care hemoglobin meters are used. However, the regulatory framework for approval, implementation and oversight of use of point-of-care hemoglobin meters is suboptimal. We assessed the diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 point-of-care hemoglobin meters, compared to a central laboratory based reference assay, in a central laboratory and a community based clinic in Durban, South Africa. Differences in performance of the point-of-car  ...[more]

Similar Datasets

| S-EPMC5685159 | biostudies-literature
| S-EPMC7529235 | biostudies-literature
| S-EPMC7282230 | biostudies-literature
| S-EPMC5505415 | biostudies-literature
| S-EPMC5120806 | biostudies-literature
| S-EPMC8617206 | biostudies-literature
| S-EPMC5224825 | biostudies-literature
| S-EPMC7019395 | biostudies-literature
| S-EPMC5481377 | biostudies-other
| S-EPMC7315854 | biostudies-literature