Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood.


ABSTRACT: There has been increasing interest in standardized and quantitative Epstein-Barr virus (EBV) DNA testing for the management of EBV disease. We evaluated the performance of the Real-Q EBV Quantification Kit (BioSewoom, Korea) in whole blood (WB). Nucleic acid extraction and real-time PCR were performed by using the MagNA Pure 96 (Roche Diagnostics, Germany) and 7500 Fast real-time PCR system (Applied Biosystems, USA), respectively. Assay sensitivity, linearity, and conversion factor were determined by using the World Health Organization international standard diluted in EBV-negative WB. We used 81 WB clinical specimens to compare performance of the Real-Q EBV Quantification Kit and artus EBV RG PCR Kit (Qiagen, Germany). The limit of detection (LOD) and limit of quantification (LOQ) for the Real-Q kit were 453 and 750 IU/mL, respectively. The conversion factor from EBV genomic copies to IU was 0.62. The linear range of the assay was from 750 to 10⁶ IU/mL. Viral load values measured with the Real-Q assay were on average 0.54 log₁₀ copies/mL higher than those measured with the artus assay. The Real-Q assay offered good analytical performance for EBV DNA quantification in WB.

SUBMITTER: Huh HJ 

PROVIDER: S-EPMC5203992 | biostudies-literature | 2017 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood.

Huh Hee Jae HJ   Park Jong Eun JE   Kim Ji Youn JY   Yun Sun Ae SA   Lee Myoung Keun MK   Lee Nam Yong NY   Kim Jong Won JW   Ki Chang Seok CS  

Annals of laboratory medicine 20170301 2


There has been increasing interest in standardized and quantitative Epstein-Barr virus (EBV) DNA testing for the management of EBV disease. We evaluated the performance of the Real-Q EBV Quantification Kit (BioSewoom, Korea) in whole blood (WB). Nucleic acid extraction and real-time PCR were performed by using the MagNA Pure 96 (Roche Diagnostics, Germany) and 7500 Fast real-time PCR system (Applied Biosystems, USA), respectively. Assay sensitivity, linearity, and conversion factor were determin  ...[more]

Similar Datasets

| S-EPMC4820504 | biostudies-literature
| S-EPMC8206016 | biostudies-literature
| S-EPMC2715772 | biostudies-literature
| S-EPMC3705264 | biostudies-literature
| S-EPMC6541347 | biostudies-literature
| S-EPMC9914079 | biostudies-literature
| S-EPMC8589012 | biostudies-literature
| S-EPMC4826252 | biostudies-literature
| S-EPMC5818632 | biostudies-literature
| S-EPMC8272980 | biostudies-literature