Reproducibility of Quantitative Brain Imaging Using a PET-Only and a Combined PET/MR System.


ABSTRACT: The purpose of this study was to test the feasibility of migrating a quantitative brain imaging protocol from a positron emission tomography (PET)-only system to an integrated PET/MR system. Potential differences in both absolute radiotracer concentration as well as in the derived kinetic parameters as a function of PET system choice have been investigated. Five healthy volunteers underwent dynamic (R)-[11C]verapamil imaging on the same day using a GE-Advance (PET-only) and a Siemens Biograph mMR system (PET/MR). PET-emission data were reconstructed using a transmission-based attenuation correction (AC) map (PET-only), whereas a standard MR-DIXON as well as a low-dose CT AC map was applied to PET/MR emission data. Kinetic modeling based on arterial blood sampling was performed using a 1-tissue-2-rate constant compartment model, yielding kinetic parameters (K1 and k2) and distribution volume (V T ). Differences for parametric values obtained in the PET-only and the PET/MR systems were analyzed using a 2-way Analysis of Variance (ANOVA). Comparison of DIXON-based AC (PET/MR) with emission data derived from the PET-only system revealed average inter-system differences of -33 ± 14% (p < 0.05) for the K1 parameter and -19 ± 9% (p < 0.05) for k2. Using a CT-based AC for PET/MR resulted in slightly lower systematic differences of -16 ± 18% for K1 and -9 ± 10% for k2. The average differences in V T were -18 ± 10% (p < 0.05) for DIXON- and -8 ± 13% for CT-based AC. Significant systematic differences were observed for kinetic parameters derived from emission data obtained from PET/MR and PET-only imaging due to different standard AC methods employed. Therefore, a transfer of imaging protocols from PET-only to PET/MR systems is not straightforward without application of proper correction methods. Clinical Trial Registration: www.clinicaltrialsregister.eu, identifier 2013-001724-19.

SUBMITTER: Lassen ML 

PROVIDER: S-EPMC5511842 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reproducibility of Quantitative Brain Imaging Using a PET-Only and a Combined PET/MR System.

Lassen Martin L ML   Muzik Otto O   Beyer Thomas T   Hacker Marcus M   Ladefoged Claes Nøhr CN   Cal-González Jacobo J   Wadsak Wolfgang W   Rausch Ivo I   Langer Oliver O   Bauer Martin M  

Frontiers in neuroscience 20170717


The purpose of this study was to test the feasibility of migrating a quantitative brain imaging protocol from a positron emission tomography (PET)-only system to an integrated PET/MR system. Potential differences in both absolute radiotracer concentration as well as in the derived kinetic parameters as a function of PET system choice have been investigated. Five healthy volunteers underwent dynamic (R)-[<sup>11</sup>C]verapamil imaging on the same day using a GE-Advance (PET-only) and a Siemens  ...[more]

Similar Datasets

| S-EPMC8356990 | biostudies-literature
| S-EPMC5453463 | biostudies-literature
| S-EPMC2917802 | biostudies-other
| S-EPMC4768229 | biostudies-literature
| S-EPMC3579175 | biostudies-literature
| S-EPMC7985095 | biostudies-literature
| S-EPMC5962598 | biostudies-literature
| S-EPMC3978416 | biostudies-other
| S-EPMC4874643 | biostudies-literature
| S-EPMC5115760 | biostudies-literature