Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor.


ABSTRACT: This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (Guardian™ Sensor 3) in the abdomen and arm.Eighty-eight subjects (14-75 years of age, mean?±?standard deviation [SD] of 42.0?±?19.1 years) with type 1 or type 2 diabetes participated in the study. Subjects wore two sensors in the abdomen that were paired with either a MiniMed™ 640G insulin pump, or an iPhone® or iPod® touch® running a glucose monitoring mobile application (Guardian Connect system) and a third sensor in the arm, which was connected to a glucose sensor recorder (GSR). Subjects were also asked to undergo in-clinic visits of 12-14?h on study days 1, 3, and 7 for frequent blood glucose sample testing using a Yellow Springs Instrument (YSI) reference.The overall mean absolute relative difference (MARD?±?SD) between abdomen sensor glucose (SG) and YSI reference values was 9.6%?±?9.0% and 9.4%?±?9.8% for the MiniMed 640G insulin pump and Guardian Connect system, respectively; and 8.7%?±?8.0% between arm SG and YSI reference values. The percentage of SG values within 20% agreement of the YSI reference value (for YSI >80?mg/dL) was 90.7% with the MiniMed 640G insulin pump, 91.8% with the Guardian Connect system, and 93.1% for GSR-connected arm sensors. Mean functional sensor life, when calibrating 3-4 times/day, was 145.9?±?39.3?h for sensors paired with the MiniMed 640G insulin pump, 146.1?±?41.6?h for sensors paired with the Guardian Connect system, and 147.6?±?40.4?h for sensors connected to the GSR. Responses to survey questions regarding sensor comfort and ease of use were favorable.The Guardian Sensor 3 glucose sensor, whether located in abdomen or the arm, provided accurate glucose readings when compared with the YSI reference and demonstrated functional life commensurate with the intended 7-day use. ClinicalTrials.gov : NCT02246582.

SUBMITTER: Christiansen MP 

PROVIDER: S-EPMC5567873 | biostudies-literature | 2017 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor.

Christiansen Mark P MP   Garg Satish K SK   Brazg Ronald R   Bode Bruce W BW   Bailey Timothy S TS   Slover Robert H RH   Sullivan Ashley A   Huang Suiying S   Shin John J   Lee Scott W SW   Kaufman Francine R FR  

Diabetes technology & therapeutics 20170712 8


<h4>Background</h4>This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (Guardian<sup>™</sup> Sensor 3) in the abdomen and arm.<h4>Methods</h4>Eighty-eight subjects (14-75 years of age, mean ± standard deviation [SD] of 42.0 ± 19.1 years) with type 1 or type 2 diabetes participated in the study. Subjects wore two sensors in the abdomen that were paired with either a MiniMed<sup>™</sup> 640G insulin pump, or an iPhone<sup>®</sup> or iPod<sup>®</sup>  ...[more]

Similar Datasets

| S-EPMC3894676 | biostudies-literature
| S-EPMC4346608 | biostudies-literature
| S-EPMC2997158 | biostudies-literature
| S-EPMC3781114 | biostudies-literature
| S-EPMC5898706 | biostudies-literature
| S-EPMC6532543 | biostudies-literature
| S-EPMC4954792 | biostudies-literature
| S-EPMC7904136 | biostudies-literature
| S-EPMC8080918 | biostudies-literature
| S-EPMC4596806 | biostudies-literature