Unknown

Dataset Information

0

Accurate measurement of chest compression depth using impulse-radio ultra-wideband sensor on a mattress.


ABSTRACT: We developed a new chest compression depth (CCD) measuring technology using radar and impulse-radio ultra-wideband (IR-UWB) sensor. This study was performed to determine its accuracy on a soft surface.Four trials, trial 1: chest compressions on the floor using an accelerometer device; trial 2: chest compressions on the floor using an IR-UWB sensor; trial 3: chest compressions on a foam mattress using an accelerometer device; trial 4: chest compressions on a foam mattress using an IR-UWB sensor, were performed in a random order. In all the trials, a cardiopulmonary resuscitation provider delivered 50 uninterrupted chest compressions to a manikin.The CCD measured by the manikin and the device were as follows: 57.42 ± 2.23 and 53.92 ± 2.92 mm, respectively in trial 1 (p < 0.001); 56.29 ± 1.96 and 54.16 ± 3.90 mm, respectively in trial 2 (p < 0.001); 55.61 ± 1.57 and 103.48 ± 10.48 mm, respectively in trial 3 (p < 0.001); 57.14 ± 3.99 and 55.51 ± 3.39 mm, respectively in trial 4 (p = 0.012). The gaps between the CCD measured by the manikin and the devices (accelerometer device vs. IR-UWB sensor) on the floor were not different (3.50 ± 2.08 mm vs. 3.15 ± 2.27 mm, respectively, p = 0.136). However, the gaps were significantly different on the foam mattress (48.53 ± 5.65 mm vs. 4.10 ± 2.47 mm, p < 0.001).The IR-UWB sensor could measure the CCD accurately both on the floor and on the foam mattress.

SUBMITTER: Yu BG 

PROVIDER: S-EPMC5576690 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accurate measurement of chest compression depth using impulse-radio ultra-wideband sensor on a mattress.

Yu Byung Gyu BG   Oh Je Hyeok JH   Kim Yeomyung Y   Kim Tae Wook TW  

PloS one 20170830 8


<h4>Objective</h4>We developed a new chest compression depth (CCD) measuring technology using radar and impulse-radio ultra-wideband (IR-UWB) sensor. This study was performed to determine its accuracy on a soft surface.<h4>Methods</h4>Four trials, trial 1: chest compressions on the floor using an accelerometer device; trial 2: chest compressions on the floor using an IR-UWB sensor; trial 3: chest compressions on a foam mattress using an accelerometer device; trial 4: chest compressions on a foam  ...[more]

Similar Datasets

| S-EPMC7093464 | biostudies-literature
| S-EPMC6115337 | biostudies-literature
| S-EPMC7769476 | biostudies-literature
| S-EPMC6695386 | biostudies-literature
| S-EPMC9416598 | biostudies-literature
| S-EPMC7147715 | biostudies-literature
| S-EPMC5167263 | biostudies-literature
| S-EPMC8792407 | biostudies-literature
| S-EPMC7312068 | biostudies-literature
| S-EPMC8575856 | biostudies-literature