Feasibility of non-contact cardiorespiratory monitoring using impulse-radio ultra-wideband radar in the neonatal intensive care unit.


ABSTRACT:

Background

Current cardiorespiratory monitoring equipment can cause injuries and infections in neonates with fragile skin. Impulse-radio ultra-wideband (IR-UWB) radar was recently demonstrated to be an effective contactless vital sign monitor in adults. The purpose of this study was to assess heart rates (HRs) and respiratory rates (RRs) in the neonatal intensive care unit (NICU) using IR-UWB radar and to evaluate its accuracy and reliability compared to conventional electrocardiography (ECG)/impedance pneumography (IPG).

Methods

The HR and RR were recorded in 34 neonates between 3 and 72 days of age during minimal movement (51 measurements in total) using IR-UWB radar (HRRd, RRRd) and ECG/IPG (HRECG, RRIPG) simultaneously. The radar signals were processed in real time using algorithms for neonates. Radar and ECG/IPG measurements were compared using concordance correlation coefficients (CCCs) and Bland-Altman plots.

Results

From the 34 neonates, 12,530 HR samples and 3,504 RR samples were measured. Both the HR and RR measured using the two methods were highly concordant when the neonates had minimal movements (CCC = 0.95 between the RRRd and RRIPG, CCC = 0.97 between the HRRd and HRECG). In the Bland-Altman plot, the mean biases were 0.17 breaths/min (95% limit of agreement [LOA] -7.0-7.3) between the RRRd and RRIPG and -0.23 bpm (95% LOA -5.3-4.8) between the HRRd and HRECG. Moreover, the agreement for the HR and RR measurements between the two modalities was consistently high regardless of neonate weight.

Conclusions

A cardiorespiratory monitor using IR-UWB radar may provide accurate non-contact HR and RR estimates without wires and electrodes for neonates in the NICU.

SUBMITTER: Lee WH 

PROVIDER: S-EPMC7769476 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Feasibility of non-contact cardiorespiratory monitoring using impulse-radio ultra-wideband radar in the neonatal intensive care unit.

Lee Won Hyuk WH   Lee Yonggu Y   Na Jae Yoon JY   Kim Seung Hyun SH   Lee Hyun Ju HJ   Lim Young-Hyo YH   Cho Seok Hyun SH   Cho Sung Ho SH   Park Hyun-Kyung HK  

PloS one 20201228 12


<h4>Background</h4>Current cardiorespiratory monitoring equipment can cause injuries and infections in neonates with fragile skin. Impulse-radio ultra-wideband (IR-UWB) radar was recently demonstrated to be an effective contactless vital sign monitor in adults. The purpose of this study was to assess heart rates (HRs) and respiratory rates (RRs) in the neonatal intensive care unit (NICU) using IR-UWB radar and to evaluate its accuracy and reliability compared to conventional electrocardiography  ...[more]

Similar Datasets

| S-EPMC7093464 | biostudies-literature
| S-EPMC6115337 | biostudies-literature
| S-EPMC6695386 | biostudies-literature
| S-EPMC7147715 | biostudies-literature
| S-EPMC6599793 | biostudies-literature
| S-EPMC5576690 | biostudies-literature
| S-EPMC9416598 | biostudies-literature
| S-EPMC7212156 | biostudies-literature
| S-EPMC5167263 | biostudies-literature
| S-EPMC5843203 | biostudies-literature