A Novel Non-contact Heart Rate Monitor Using Impulse-Radio Ultra-Wideband (IR-UWB) Radar Technology.


ABSTRACT: We discovered that impulse-radio ultra-wideband (IR-UWB) radar could recognize cardiac motions in a non-contact fashion. Therefore, we measured the heart rate (HR) and rhythms using an IR-UWB radar sensor and evaluated the validity and reliability of the measurements in comparison to electrocardiography. The heart beats were measured in 6 healthy volunteers (18 samples) with normal sinus rhythm (NSR) and 16 patients (36 samples) with atrial fibrillation (AF) using both an IR-UWB radar sensor and electrocardiography simultaneously. The participants hold their breath for 20?seconds during the data acquisition. In subjects with NSR, there was excellent agreement of HR (intraclass correlation coefficient [ICC] 0.856), average R-R interval (ICC 0.997) and individual R-R intervals between the two methods (ICC 0.803). In subjects with AF, HR (ICC 0.871) and average R-R interval (ICC 0.925) from the radar sensor also agreed well with those from electrocardiography, though there was a small disagreement in the individual R-R intervals between the two methods (ICC 0.697). The rhythms computed by the signal-processing algorithm showed good agreement between the two methods (Cohen's Kappa 0.922). The IR-UWB radar sensor is precise and accurate for assessing HR and rhythms in a non-contact fashion.

SUBMITTER: Lee Y 

PROVIDER: S-EPMC6115337 | biostudies-literature | 2018 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

A Novel Non-contact Heart Rate Monitor Using Impulse-Radio Ultra-Wideband (IR-UWB) Radar Technology.

Lee Yonggu Y   Park Jun-Young JY   Choi Yeon-Woo YW   Park Hyun-Kyung HK   Cho Seok-Hyun SH   Cho Sung Ho SH   Lim Young-Hyo YH  

Scientific reports 20180829 1


We discovered that impulse-radio ultra-wideband (IR-UWB) radar could recognize cardiac motions in a non-contact fashion. Therefore, we measured the heart rate (HR) and rhythms using an IR-UWB radar sensor and evaluated the validity and reliability of the measurements in comparison to electrocardiography. The heart beats were measured in 6 healthy volunteers (18 samples) with normal sinus rhythm (NSR) and 16 patients (36 samples) with atrial fibrillation (AF) using both an IR-UWB radar sensor and  ...[more]

Similar Datasets

| S-EPMC7093464 | biostudies-literature
| S-EPMC7769476 | biostudies-literature
| S-EPMC6695386 | biostudies-literature
| S-EPMC7147715 | biostudies-literature
| S-EPMC6599793 | biostudies-literature
| S-EPMC5576690 | biostudies-literature
| S-EPMC8100299 | biostudies-literature
| S-EPMC8041886 | biostudies-literature
| S-EPMC9416598 | biostudies-literature
| S-EPMC9392064 | biostudies-literature