A novel optimization framework for VMAT with dynamic gantry couch rotation.


ABSTRACT: Existing volumetric modulated arc therapy (VMAT) optimization using coplanar arcs is highly efficient but usually dosimetrically inferior to intensity modulated radiation therapy (IMRT) with optimized non-coplanar beams. To achieve both dosimetric quality and delivery efficiency, we proposed in this study, a novel integrated optimization method for non-coplanar VMAT (4?VMAT). 4?VMAT with direct aperture optimization (DAO) was achieved by utilizing a least square dose fidelity objective, along with an anisotropic total variation term for regularizing the fluence smoothness, a single segment term for imposing simple apertures, and a group sparsity term for selecting beam angles. Continuous gantry/couch angle trajectories were selected using the Dijkstra's algorithm, where the edge and node costs were determined based on the maximal gantry rotation speed and the estimated fluence map at the current iteration, respectively. The couch-gantry-patient collision space was calculated based on actual machine geometry and a human subject 3D surface. Beams leading to collision are excluded from the DAO and beam trajectory selection (BTS). An alternating optimization strategy was implemented to solve the integrated DAO and BTS problem. The feasibility of 4?VMAT using one full-arc or two full-arcs was tested on nine patients with brain, lung, or prostate cancer. The plan was compared against a coplanar VMAT (2?VMAT) plan using one additional arc and collimator rotation. Compared to 2?VMAT, 4?VMAT reduced the average maximum and mean organs-at-risk dose by 9.63% and 3.08% of the prescription dose with the same target coverage. R50 was reduced by 23.0%. Maximum doses to the dose limiting organs, such as the brainstem, the major vessels, and the proximal bronchus, were reduced by 8.1 Gy (64.8%), 16.3 Gy (41.5%), and 19.83 Gy (55.5%), respectively. The novel 4?VMAT approach affords efficient delivery of non-coplanar arc trajectories that lead to dosimetric improvements compared with coplanar VMAT using more arcs.

SUBMITTER: Lyu Q 

PROVIDER: S-EPMC6075870 | biostudies-literature | 2018 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

A novel optimization framework for VMAT with dynamic gantry couch rotation.

Lyu Qihui Q   Yu Victoria Y VY   Ruan Dan D   Neph Ryan R   O'Connor Daniel D   Sheng Ke K  

Physics in medicine and biology 20180613 12


Existing volumetric modulated arc therapy (VMAT) optimization using coplanar arcs is highly efficient but usually dosimetrically inferior to intensity modulated radiation therapy (IMRT) with optimized non-coplanar beams. To achieve both dosimetric quality and delivery efficiency, we proposed in this study, a novel integrated optimization method for non-coplanar VMAT (4πVMAT). 4πVMAT with direct aperture optimization (DAO) was achieved by utilizing a least square dose fidelity objective, along wi  ...[more]

Similar Datasets

| S-EPMC5713922 | biostudies-other
| S-EPMC5711247 | biostudies-literature
| S-EPMC5711073 | biostudies-literature
| S-EPMC5690937 | biostudies-literature
| S-EPMC8374176 | biostudies-literature
| S-EPMC6731354 | biostudies-literature
| S-EPMC8861408 | biostudies-literature
| S-EPMC5514241 | biostudies-other
| S-EPMC8791528 | biostudies-literature
| S-EPMC7182202 | biostudies-literature