Fukuda Atsushi A Lin Pei-Jan P PJ Matsubara Kosuke K Miyati Tosiaki T
Journal of applied clinical medical physics 20140908 5
The purpose of this study was to develop and evaluate a simple method to assess gantry rotation overrun in a single axial CT scanning. The exposure time in the axial scanning was measured at selected nominal rotation times (400, 700, and 1000 ms) using a solid-state detector, the RTI's CT dose profiler (CTDP). CTDP was placed at the isocenter and the radiation dose rate signal (profile) was recorded. Subsequently, the full width of this profile was determined as the exposure time (Taxial). Next, ...[more]