Performance Evaluation of G8, a High-Sensitivity Benchtop Preclinical PET/CT Tomograph.


ABSTRACT: G8 is a benchtop integrated PET/CT scanner dedicated to high-sensitivity and high-resolution imaging of mice. This work characterizes its National Electrical Manufacturers Association NU 4-2008 performance where applicable and also assesses the basic imaging performance of the CT subsystem. Methods: The PET subsystem in G8 consists of 4 flat-panel detectors arranged in a boxlike geometry. Each panel consists of 2 modules of a 26 × 26 pixelated bismuth germanate scintillator array with individual crystals measuring 1.75 × 1.75 × 7.2 mm. The crystal arrays are coupled to multichannel photomultiplier tubes via a tapered, pixelated glass lightguide. A cone-beam CT scanner consisting of a MicroFocus x-ray source and a complementary metal oxide semiconductor detector provides anatomic information. Sensitivity, spatial resolution, energy resolution, scatter fraction, count-rate performance, and the capability of performing phantom and mouse imaging were evaluated for the PET subsystem. Noise, dose level, contrast, and resolution were evaluated for the CT subsystem. Results: With an energy window of 350-650 keV, the peak sensitivity was 9.0% near the center of the field of view. The crystal energy resolution ranged from 15.0% to 69.6% in full width at half maximum (FWHM), with a mean of 19.3% ± 3.7%. The average intrinsic spatial resolution was 1.30 and 1.38 mm FWHM in the transverse and axial directions, respectively. The maximum-likelihood expectation maximization reconstructed image of a point source in air averaged 0.81 ± 0.11 mm FWHM. The peak noise-equivalent count rate for the mouse-sized phantom was 44 kcps for a total activity of 2.9 MBq (78 ?Ci), and the scatter fraction was 11%. For the CT subsystem, the value of the modulation transfer function at 10% was 2.05 cycles/mm. Conclusion: The overall performance demonstrates that the G8 can produce high-quality images for molecular imaging-based biomedical research.

SUBMITTER: Gu Z 

PROVIDER: S-EPMC6354226 | biostudies-literature | 2019 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance Evaluation of G8, a High-Sensitivity Benchtop Preclinical PET/CT Tomograph.

Gu Zheng Z   Taschereau Richard R   Vu Nam T NT   Prout David L DL   Silverman Robert W RW   Lee Jason T JT   Chatziioannou Arion F AF  

Journal of nuclear medicine : official publication, Society of Nuclear Medicine 20180614 1


G8 is a benchtop integrated PET/CT scanner dedicated to high-sensitivity and high-resolution imaging of mice. This work characterizes its National Electrical Manufacturers Association NU 4-2008 performance where applicable and also assesses the basic imaging performance of the CT subsystem. <b>Methods:</b> The PET subsystem in G8 consists of 4 flat-panel detectors arranged in a boxlike geometry. Each panel consists of 2 modules of a 26 × 26 pixelated bismuth germanate scintillator array with ind  ...[more]

Similar Datasets

| S-EPMC4009355 | biostudies-literature
| S-EPMC7728905 | biostudies-literature
| S-EPMC8455202 | biostudies-literature
| S-EPMC9616809 | biostudies-literature
| S-EPMC6509903 | biostudies-other
| S-EPMC8263427 | biostudies-literature
| S-EPMC10994244 | biostudies-literature
| S-EPMC7237803 | biostudies-literature
| S-EPMC4758607 | biostudies-literature
| S-EPMC5745571 | biostudies-literature