ABSTRACT: Purpose
In ultrashort echo time (UTE) imaging, fat suppression can improve short T2 * contrast but can also reduce short T2 * signals. The conventional two-point Dixon (2p-Dixon) method does not perform well due to short T2 * decay. In this study, we propose a new method to suppress fat for high contrast UTE imaging of short T2 tissues, utilizing a single-point Dixon (1p-Dixon) method.Methods
The proposed method utilizes dual-echo UTE imaging, where UTE is followed by the second TE, chosen flexibly. Fat is estimated by applying a 1p-Dixon method to the non-UTE image after correction of phase errors, which is used to suppress fat in the UTE image. In vivo ankle and knee imaging were performed at 3 T to evaluate the proposed method.Result
It was observed that fat and water signals in tendons were misestimated by the 2p-Dixon method due to signal decay, while the 1p-Dixon method showed reliable fat and water separation not affected by the short T2 * signal decay. Compared with the conventional chemical shift based fat saturation technique, the 1p-Dixon based approach showed much stronger signal intensities in the Achilles, quadriceps, and patellar tendons, with significantly improved contrast to noise ratios (CNRs) of 11.8 ± 2.2, 16.0 ± 1.6, and 26.8 ± 1.3 with the 1p-Dixon method and 0.6 ± 0.2, 4.6 ± 1.0, and 17.5 ± 1.4 with regular fat saturation, respectively.Conclusion
The proposed 1p-Dixon based fat suppression allows more flexible selection of imaging parameters and more accurate fat and water separation over the conventional 2p-Dixon in UTE imaging. Moreover, the proposed method provides much improved CNR for short T2 tissues over the conventional fat saturation method.
SUBMITTER: Jang H
PROVIDER: S-EPMC6476675 | biostudies-literature | 2019 May
REPOSITORIES: biostudies-literature
Jang Hyungseok H Carl Michael M Ma Yajun Y Jerban Saeed S Guo Tan T Zhao Wei W Chang Eric Y EY Du Jiang J
NMR in biomedicine 20190215 5
<h4>Purpose</h4>In ultrashort echo time (UTE) imaging, fat suppression can improve short T<sub>2</sub> * contrast but can also reduce short T<sub>2</sub> * signals. The conventional two-point Dixon (2p-Dixon) method does not perform well due to short T<sub>2</sub> * decay. In this study, we propose a new method to suppress fat for high contrast UTE imaging of short T<sub>2</sub> tissues, utilizing a single-point Dixon (1p-Dixon) method.<h4>Methods</h4>The proposed method utilizes dual-echo UTE i ...[more]