ABSTRACT: Purpose
In ultrashort echo time (UTE) imaging, fat suppression can improve short T2 * contrast but can also reduce short T2 * signals. The conventional two-point Dixon (2p-Dixon) method does not perform well due to short T2 * decay. In this study, we propose a new method to suppress fat for high contrast UTE imaging of short T2 tissues, utilizing a single-point Dixon (1p-Dixon) method.Methods
The proposed method utilizes dual-echo UTE imaging, where UTE is followed by the second TE, chosen flexibly. Fat is estimated by applying a 1p-Dixon method to the non-UTE image after correction of phase errors, which is used to suppress fat in the UTE image. In vivo ankle and knee imaging were performed at 3 T to evaluate the proposed method.