Unknown

Dataset Information

0

Laboratory Evaluation of Linearity, Repeatability, and Hematocrit Interference With an Internet-Enabled Blood Glucose Meter.


ABSTRACT: BACKGROUND:In recent clinical trials, use of the MyGlucoHealth blood glucose meter (BGM) and electronic diary was associated with an unusual reporting pattern of glycemic data and hypoglycemic events. Therefore, the performance of representative BGMs used by the patients was investigated to assess repeatability, linearity, and hematocrit interference in accordance with regulatory guidelines. METHOD:Ten devices and 6 strip lots were selected using standard randomization and repeatability procedures. Venous heparinized blood was drawn from healthy subjects, immediately aliquoted and adjusted to 5 target blood glucose (BG) ranges for the repeatability and 11 BG concentrations for the linearity tests. For the hematocrit interference test, each sample within 5 target BG ranges was split into 5 aliquots and adjusted to hematocrit levels across the acceptance range. YSI 2300 STAT Plus was used as the laboratory reference method in all experiments. RESULTS:Measurement repeatability or precision was acceptable across the target BG ranges for all devices and strip lots with coefficient of variation (CV) between 3.4-9.7% (mean: 5.7%). Linearity was shown by a correlation coefficient of .991; however, a positive bias was seen for BG <100 mg/dL (86% measurements did not meet ISO15197:2015 acceptance criteria). Significant hematocrit interference (up to 20%) was observed for BG >100 mg/dL (ISO15197:2015 acceptance criteria: ±10%), while the results were acceptable for BG <100 mg/dL. CONCLUSIONS:The BGM met repeatability requirements but demonstrated a significant measurement bias in the low BG range. In addition, it failed the ISO15197:2015 criteria for hematocrit interference.

SUBMITTER: Demircik F 

PROVIDER: S-EPMC6501519 | biostudies-literature | 2019 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Laboratory Evaluation of Linearity, Repeatability, and Hematocrit Interference With an Internet-Enabled Blood Glucose Meter.

Demircik Filiz F   Kirsch Valeria V   Ramljak Sanja S   Vogg Mario M   Pfützner Anke H AH   Pfützner Andreas A  

Journal of diabetes science and technology 20190411 3


<h4>Background</h4>In recent clinical trials, use of the MyGlucoHealth blood glucose meter (BGM) and electronic diary was associated with an unusual reporting pattern of glycemic data and hypoglycemic events. Therefore, the performance of representative BGMs used by the patients was investigated to assess repeatability, linearity, and hematocrit interference in accordance with regulatory guidelines.<h4>Method</h4>Ten devices and 6 strip lots were selected using standard randomization and repeata  ...[more]

Similar Datasets

| S-EPMC6501533 | biostudies-literature
| S-EPMC6803884 | biostudies-literature
| S-EPMC2494647 | biostudies-literature
| S-EPMC5851235 | biostudies-literature
| S-EPMC4764205 | biostudies-literature
| S-EPMC5224825 | biostudies-literature
| S-EPMC7090327 | biostudies-literature
| S-EPMC7463700 | biostudies-literature
| S-EPMC5934745 | biostudies-literature
| S-EPMC8660689 | biostudies-literature