Non-contact respiration monitoring using impulse radio ultrawideband radar in neonates.


ABSTRACT: Vital sign monitoring in neonates requires adhesive electrodes, which often damage fragile newborn skin. Because impulse radio ultrawideband (IR-UWB) radar has been reported to recognize chest movement without contact in adult humans, IR-UWB may be used to measure respiratory rates (RRs) in a non-contact fashion. We investigated the feasibility of radar sensors for respiration monitoring in neonates without any respiratory support to compare the accuracy and reliability of radar measurements with those of conventional impedance pneumography measurements. In the neonatal intensive care unit, RRs were measured using radar (RRRd) and impedance pneumography (RRIP) simultaneously. The neonatal voluntary movements were measured using the radar sensor and categorized into three levels (low [M0], intermediate [M1] and high [M2]). RRRd highly agreed with RRIP (r = 0.90; intraclass correlation coefficient [ICC] = 0.846 [0.835-0.856]). For the M0 movement, there was good agreement between RRRd and RRIP (ICC = 0.893; mean bias -0.15 [limits of agreement (LOA) -9.6 to 10.0]). However, the agreement was slightly lower for the M1 (ICC = 0.833; mean bias = 0.95 [LOA -11.4 to 13.3]) and M2 (ICC = 0.749; mean bias = 3.04 [LOA -9.30 to 15.4]) movements than for the M0 movement. In conclusion, IR-UWB radar can provide accurate and reliable estimates of RR in neonates in a non-contact fashion. The performance of radar measurements could be affected by neonate movement.

SUBMITTER: Kim JD 

PROVIDER: S-EPMC6599793 | biostudies-literature | 2019 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Non-contact respiration monitoring using impulse radio ultrawideband radar in neonates.

Kim Jong Deok JD   Lee Won Hyuk WH   Lee Yonggu Y   Lee Hyun Ju HJ   Cha Teahyen T   Kim Seung Hyun SH   Song Ki-Min KM   Lim Young-Hyo YH   Cho Seok Hyun SH   Cho Sung Ho SH   Park Hyun-Kyung HK  

Royal Society open science 20190605 6


Vital sign monitoring in neonates requires adhesive electrodes, which often damage fragile newborn skin. Because impulse radio ultrawideband (IR-UWB) radar has been reported to recognize chest movement without contact in adult humans, IR-UWB may be used to measure respiratory rates (RRs) in a non-contact fashion. We investigated the feasibility of radar sensors for respiration monitoring in neonates without any respiratory support to compare the accuracy and reliability of radar measurements wit  ...[more]

Similar Datasets

| S-EPMC7093464 | biostudies-literature
| S-EPMC7769476 | biostudies-literature
| S-EPMC6115337 | biostudies-literature
| S-EPMC6695386 | biostudies-literature
| S-EPMC8688557 | biostudies-literature
| S-EPMC5638502 | biostudies-literature
| S-EPMC10452166 | biostudies-literature
| S-EPMC9586033 | biostudies-literature
| S-EPMC8041886 | biostudies-literature
| S-EPMC8157066 | biostudies-literature