ABSTRACT: Background
This study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition. The method should offer high resolution, accuracy, precision and reproducibility.Methods
The proposed reconstruction estimates myocardial parameter maps directly from undersampled k-space which is continuously measured by IR radial FLASH with a 4?s breathhold and retrospectively sorted based on a cardiac trigger signal. Joint sparsity constraints are imposed on the parameter maps to further improve T1 precision. Validations involved studies of an experimental phantom and 8 healthy adult subjects.Results
In comparison to an IR spin-echo reference method, phantom experiments with T1 values ranging from 300 to 1500?ms revealed good accuracy and precision at simulated heart rates between 40 and 100?bpm. In vivo T1 maps achieved better precision and qualitatively better preservation of image features for the proposed method than a real-time CMR approach followed by pixelwise fitting. Apart from good inter-observer reproducibility (0.6% of the mean), in vivo results confirmed good intra-subject reproducibility (1.05% of the mean for intra-scan and 1.17, 1.51% of the means for the two inter-scans, respectively) of the proposed method.Conclusion
Model-based reconstructions with sparsity constraints allow for single-shot myocardial T1 maps with high spatial resolution, accuracy, precision and reproducibility within a 4?s breathhold. Clinical trials are warranted.
SUBMITTER: Wang X
PROVIDER: S-EPMC6751613 | biostudies-literature | 2019 Sep
REPOSITORIES: biostudies-literature
Wang Xiaoqing X Kohler Florian F Unterberg-Buchwald Christina C Lotz Joachim J Frahm Jens J Uecker Martin M
Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance 20190919 1
<h4>Background</h4>This study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition. The method should offer high resolution, accuracy, precision and reproducibility.<h4>Methods</h4>The proposed reconstruction estimates myocardial parameter maps directly from undersampled k-space which is continuously measured by IR radial FLASH wit ...[more]