Optimization of b-value schemes for estimation of the diffusion coefficient and the perfusion fraction with segmented intravoxel incoherent motion model fitting.


ABSTRACT:

Purpose

Intravoxel incoherent motion (IVIM) modeling for estimation of the diffusion coefficient (D) and perfusion fraction (f) is increasingly popular, but no consensus on standard protocols exists. This study provides a framework for optimization of b-value schemes for reduced estimation uncertainty of D and f from segmented model fitting.

Theory

Analytical expressions for uncertainties of D and f from segmented model fitting were derived as Cramer-Rao lower bounds (CRLBs).

Methods

Optimized b-value schemes were obtained for 3 to 12 acquisitions and in the limit of infinitely many acquisitions through constrained minimization of the CRLBs, with b-values constrained to be 0 or 200 to 800 s/mm2 . The optimized b-value scheme with eight acquisitions was compared with b-values linearly distributed in the allowed range using simulations and in vivo liver data from seven healthy volunteers.

Results

All optimized b-value schemes contained exactly three unique b-values regardless of the total number of acquisitions (0, 200, and 800 s/mm2 ) with repeated acquisitions distributed approximately as 1:2:2. Compared with linearly distributed b-values, the variability of estimates of D and f was reduced by approximately 30% as seen both in simulations and in repeated in vivo measurements.

Conclusion

The uncertainty of IVIM D and f estimates can be reduced by the use of optimized b-value schemes.

SUBMITTER: Jalnefjord O 

PROVIDER: S-EPMC6772171 | biostudies-literature | 2019 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Optimization of b-value schemes for estimation of the diffusion coefficient and the perfusion fraction with segmented intravoxel incoherent motion model fitting.

Jalnefjord Oscar O   Montelius Mikael M   Starck Göran G   Ljungberg Maria M  

Magnetic resonance in medicine 20190531 4


<h4>Purpose</h4>Intravoxel incoherent motion (IVIM) modeling for estimation of the diffusion coefficient (D) and perfusion fraction (f) is increasingly popular, but no consensus on standard protocols exists. This study provides a framework for optimization of b-value schemes for reduced estimation uncertainty of D and f from segmented model fitting.<h4>Theory</h4>Analytical expressions for uncertainties of D and f from segmented model fitting were derived as Cramer-Rao lower bounds (CRLBs).<h4>M  ...[more]

Similar Datasets

| S-EPMC5674133 | biostudies-literature
| S-EPMC5412931 | biostudies-literature
| S-EPMC4520473 | biostudies-literature
| S-EPMC7453742 | biostudies-literature
| S-EPMC6886883 | biostudies-literature
| S-EPMC3758329 | biostudies-literature
| S-EPMC7206040 | biostudies-literature
| S-EPMC5926247 | biostudies-literature
| S-EPMC5832256 | biostudies-literature
| S-EPMC6001424 | biostudies-literature