Unknown

Dataset Information

0

Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer.


ABSTRACT:

Purpose

To compare biometry and prediction of postoperative refractive outcomes obtained by two swept-source optical coherence tomography (SS-OCT) biometers (IOLMaster 700 and Argos), and a partial coherence interferometry (IOLMaster ver 5.4).

Methods

Biometric values were measured using two SS-OCT and PCI device and evaluated against one another. Predictive errors were compared at one month after cataract surgery.

Results

One hundred forty six eyes were considered. Axial length (AXL) measurements were not successful in 3 eyes measured by IOLMaster 700 and Argos devices, and in 17 eyes measured by IOLMaster ver. 5.4 devices. AXL as measured by Argos showed a tendency to be shorter in long eyes with AXL more than 26.0 mm (p < .001) and to be longer in short eyes with AXL less than 22.5 mm (p = .005). Anterior chamber depth as measured by IOLMaster ver. 5.4 was longer than that measured by the other two SS-OCT devices (vs. IOLMaster 700: p = .003; vs. Argos: p = .006). White-to-white diameter measured using Argos was significantly different measurements obtained using two IOLMaster (p < .001, respectively). The mean absolute postoperative prediction errors were 0.41 ± 0.31 diopters (D), 0.42 ± 0.32 D, and 0.35 ± 0.30 D for IOLMaster ver. 5.4, IOLMaster 700, and Argos, respectively.

Conclusion

The ocular biometric measurements using three devices showed high agreement. AXL measured by Argos showed a significant difference compared with the measurements from two IOLMaster. ACD was highly correlated between two SS-OCT devices except IOLMaster ver 5.4. LT and CCT values between IOLMaster 700 and Argos were different significantly. SS-OCT devices demonstrated a superior ability to successfully perform measurements compared with PCI device.

SUBMITTER: Yang CM 

PROVIDER: S-EPMC6788676 | biostudies-literature | 2019

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer.

Yang Chan Min CM   Lim Dong Hui DH   Kim Hyo Jeong HJ   Chung Tae-Young TY  

PloS one 20191011 10


<h4>Purpose</h4>To compare biometry and prediction of postoperative refractive outcomes obtained by two swept-source optical coherence tomography (SS-OCT) biometers (IOLMaster 700 and Argos), and a partial coherence interferometry (IOLMaster ver 5.4).<h4>Methods</h4>Biometric values were measured using two SS-OCT and PCI device and evaluated against one another. Predictive errors were compared at one month after cataract surgery.<h4>Results</h4>One hundred forty six eyes were considered. Axial l  ...[more]

Similar Datasets

| S-EPMC9388892 | biostudies-literature
| S-EPMC8947380 | biostudies-literature
| S-EPMC5918161 | biostudies-literature
| S-EPMC8253773 | biostudies-literature
| S-EPMC6137182 | biostudies-literature
| S-EPMC7775111 | biostudies-literature
| S-EPMC3296548 | biostudies-literature
| S-EPMC7066125 | biostudies-literature
| S-EPMC5440072 | biostudies-literature
| S-EPMC6307737 | biostudies-literature