Comparison study of the axial length measured using the new swept-source optical coherence tomography ANTERION and the partial coherence interferometry IOL Master.


ABSTRACT:

Purpose

To compare a biometer using swept-source optical coherence tomography (SS-OCT) with a partial coherence interferometry (PCI)-based biometer in measurements of two ocular biometry parameters, i.e., the axial length and anterior cornea curvature.

Methods

We compared the two biometers SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOL Master, Carl Zeiss Meditec, Jena, Germany) in terms of the axial length (AL) and corneal curvature (K) measurements of 175 eyes. Paired t-tests were used to compare the two biometers. Agreement between the biometers was evaluated using the Bland-Altman method.

Results

The mean age was 36.0 ± 25.6 years (range: 5 to 85 years). The mean axial length was 24.42 ± 0.13 mm for SS-OCT and 24.45 ± 0.14 mm for PCI. The mean corneal curvature was significantly different between the two biometry in flat K (K1) but not in steep K (K2). The limit of agreement was -0.15 to 0.21 in the axial length, -1.18 to 0.83 in K1, and -1.06 to 0.95 in K2. All above ocular biometric measurements between SS-OCT and PCI correlated significantly (Pearson's correlation, p<0.001).

Conclusions

The axial length measured using SS-OCT is useful in clinical practice. It shows a good correlation and agreement with that measured using PCI. However, the axial length and corneal curvature measured using SS-OCT cannot be used interchangeably with that measured using PCI in clinical practice.

SUBMITTER: Kim KY 

PROVIDER: S-EPMC7775111 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison study of the axial length measured using the new swept-source optical coherence tomography ANTERION and the partial coherence interferometry IOL Master.

Kim Kook Young KY   Choi Gon Soo GS   Kang Min Seok MS   Kim Ungsoo Samuel US  

PloS one 20201231 12


<h4>Purpose</h4>To compare a biometer using swept-source optical coherence tomography (SS-OCT) with a partial coherence interferometry (PCI)-based biometer in measurements of two ocular biometry parameters, i.e., the axial length and anterior cornea curvature.<h4>Methods</h4>We compared the two biometers SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOL Master, Carl Zeiss Meditec, Jena, Germany) in terms of the axial length (AL) and corneal curvature (K) measuremen  ...[more]

Similar Datasets

| S-EPMC6307737 | biostudies-literature
| S-EPMC6137182 | biostudies-literature
| S-EPMC7872294 | biostudies-literature
| S-EPMC5918161 | biostudies-literature
| S-EPMC8096100 | biostudies-literature
| S-EPMC7066125 | biostudies-literature
| S-EPMC6788676 | biostudies-literature
| S-EPMC9388892 | biostudies-literature
| S-EPMC9403934 | biostudies-literature
| S-EPMC7556480 | biostudies-literature