Unknown

Dataset Information

0

Performance of InBios ZIKV Detect™ 2.0 IgM Capture ELISA in two reference laboratories compared to the original ZIKV Detect™ IgM Capture ELISA.


ABSTRACT: ZIKV Detect™ 2.0 IgM Capture ELISA (InBios International, Seattle, WA) recently replaced the ZIKV Detect™ IgM Capture ELISA and a number of significant changes have been made to the original version. This study compares data generated from the ZIKV Detect™ 2.0 IgM Capture ELISA, to data generated using the original version of the kit. The same sample sets were used in this comparison, and reference test results for these samples were used to assess sensitivity, specificity, accuracy and concordance of results across two laboratories. Average sensitivity increased from 90.4% to 92.5% with the updated kit where the increase was not statistically different, and specificity increased from 79.5% to 97.4%, a statistically-significant difference. Accuracy of the ZIKV Detect™ 2.0 IgM Capture ELISA was 89% compared to 63.9% for the original version of the kit, and agreement across the laboratories increased from 79.5% to 97.4%. With secondary dengue virus infections, specificity increased from 9.3% to 82.6% with the updated kit, primarily due to the change in interpretation criteria that no longer includes "Possible Zika positive."

SUBMITTER: Basile AJ 

PROVIDER: S-EPMC7176032 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC9611129 | biostudies-literature
| S-EPMC2615055 | biostudies-literature
| S-EPMC4625539 | biostudies-literature
| S-EPMC6000977 | biostudies-literature
| S-EPMC6640698 | biostudies-literature
| S-EPMC4472754 | biostudies-literature
| S-EPMC5977246 | biostudies-literature
| S-EPMC6302594 | biostudies-literature
| S-EPMC5666375 | biostudies-literature
| PRJNA266138 | ENA